additional, not regulated by IEC (their change does not affect the
settings in accordance with the requirements of IEC):
On subsequent switching operation, the Oil tester operates with the
latest saved test parameters;
b) set the power switch
in position "І". Green LED indicator
must
light up in green. The LCD display will show one of these screenshots:
1) in case of absence of
mounted cell;
2) in case of opened lid on test
compartment;
3) both 1 and 2;
c) mount the cell with insulating liquid dielectric on high-voltage termi-
nals inside the test compartment. Immerse activating rod to the bottom of the
cell. Close test compartment lid. The LCD display will indicate following:
;
*
- for models UIM-90 (221/В); UIM-90 (221/ВР)
**
- for models UIM-90 (221/P); UIM-90 (221/ВР)
P
- icon of built-in thermal printer, displayed on models UIM-90 (221/P); UIM-90 (221/BP)
- icon to connect the Oil tester to the network when the built-in battery is available
- charge level icon if there is a built-in battery
Stirring time Ts, sec
10
Stirring speed
⅔
Bluetooth
*
OFF
Printing
**
ON
Type of electrodes
PART. SPHER.