Instructions for use (rev 05/18)
e.ON SINTER
UGIN DENTAIRE
38
NOTE:
High translucency zirconia are very sensitives. To avoid and prevent crossed discoloration with
other zirconia brand or type, user must do regularly clean program (N° 40).
If liquid shades are used with staining technic, prosthetics units must be dry and preheated with
an infra red lamp or into an oven to prevent pollution into the eON Sinter, in following zirconia
and shades manufacturer recommandations.
White clusters can appear on the rods (heating elements) when differents brands or types of
zirconia are used. In this case user must clean mufle and rods with a special zirconia cleaner like
Nacera Clean (registered brand) with clean program N° 40.
6.6.2.
Exemple of programs
Next, and as an example, four sintering curves:
Sintering curve No. 1
(1 to 3 unit or bridges)
Sintering curve No. 2
(4 to 5 unit or bridges)
Stage
Temperature
(°C)
Time
(minutes)
Stage
Temperature
(°C)
Time
(minutes)
1
C1: 0
T1: 30
1
C1: 0
T1: 30
2
C2: 300
T2: 30
2
C2: 300
T2: 60
3
C3: 1000
T3: 120
3
C3: 1000
T3: 180
4
C4: 1480
T4: 60
4
C4: 1480
T4: 90
5
C5: 1480
T5: 60
5
C5: 1480
T5: 60
6
C6: 800
T6: -121
6
C6: 800
T6: -121
Sintering curve No. 3
(6 to 8 unit or bridges)
Sintering curve No. 4
(9 and above unit or bridges)
Stage
Temperature
(°C)
Time
(minutes)
Stage
Temperature
(°C)
Time
(minutes)
1
C1: 0
T1: 30
1
C1: 0
T1: 30
2
C2: 300
T2: 60
2
C2: 300
T2: 90
3
C3: 1000
T3: 240
3
C3: 1000
T3: 360
4
C4: 1480
T4: 90
4
C4: 1480
T4: 90
5
C5: 1480
T5: 90
5
C5: 1480
T5: 90
6
C6: 800
T6: -121
6
C6: 800
T6: -121
Содержание e.ON SINTER