background image

Electrical

17

Measurement range 

J-Type: -346˚ to +2192˚F (-210˚ to +1200˚C)

 

K-Type: -328˚ to +2498˚F (-200˚ to +1370˚C)

 

T-Type: -418˚ to +752˚F (-250˚ to +400˚C)

 

E-Type: -238˚ to +1832˚F (-150˚ to +1000˚C)

Display Resolution 

0.1˚F/˚C < 1000˚

 1.0˚F/˚C 

 1000˚

Measurement 

J, K, T, and E-Type; ±[0.1% +1.4˚F (0.8˚C)]

Accuracy 

[Below -148˚F (-100˚C): add 0.2% of reading

 

for J, K, and E-Type; and 0.5% of reading for 

 T-Type]
Temperature 

0.01% of reading 0.1˚F per ˚F(+0.05˚C per ˚C) 

Coefficient 

for <+64˚F (+18˚C) or +82˚F (+28˚C) 

 

[Below -148˚F (-100˚C): add 0.05% of reading

 

for J, K, and E-type; and 0.1% of reading for 

 T-type]
Real Time 

About 1 second per day

Clock Tolerance
Maximum Differential  1V (maximum voltage difference between
Common Mode  

any pair of inputs)

Voltage
Temperature Scale 

ITS-90 (International Temperature Scale  

 

of 1990)

Applicable Standards  N.I.S.T. Monograph 175 revised to ITS-90
Accuracy is specified for ambient temperatures between 64˚F (18˚C) 
and 82˚F (28˚C) for a period of 1 year. The above specifications do not 
include thermocouple error.

16

Specifications

Environmental

General

Thermocouple

Operating Temperature 

14˚ to 122˚F (-10˚ to 50˚C)

Storage Temperature 

-40˚ to 140˚F (-40˚ to 60˚C)

Humidity 

Non condensing <50˚F (10˚C)

 

85% RH: 50˚ to 86˚F (10˚ to 30˚C)

 

70% RH: 86˚ to 104˚F (30˚ to 40˚C)

 

45% RH: 104˚ to 122˚F (40˚ to 50˚C)

Altitude 

Operating - up to 200 m

 

Storage - 10000 m 

Dimension 

18.3 (H) x 9.4 (W) x 4.3 (D) cm

 

(7.20” x 3.70” x 1.70)

Weight 

Approx. 460 g (16.2 oz)

Battery 

2 x LR03 (AAA) type 1.5V

Certification CE
Safety 

IEC 1010-1 (2001), UL 3111-1 (6, 1994),

 

EN 61010-1 (2001), CSA C22.2  

 

No. 1010.1 (1992)

CAT I 

OVERVOLTAGE (Installation) CATEGORY I,  

 

Pollution Degree 2 per IEC 1010-1

Immersion (30 min) 

IP67 according to IEC 60529

& Dust proof

Instrument  Part # 

Type 

Specifications

DT301 

ATT19 

J-Type 

(-40˚- 950˚F/-40˚-510˚C) 48”  

    
DT302 

ATT39 

K-Type 

-40˚ to +900˚F (-40˚ to +480˚C)

  

 

 

Accuracy: ±2.0˚F (±1.1˚C)

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание DT301

Страница 1: ...DT301 DT302 Digital Temperature Logger INSTRUCTION MANUAL www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...the proper thermocouple type you are using Be sure the thermocouple you use can withstand the temperature extreme it may be exposed to in your service task Properly maintain your thermometer and cali...

Страница 3: ...C F K Temperature units 11 Secondary Display Internal temperature of the thermometer MEMORY MIN MAX AVG and T1 or T2 DT302 reading Shows the memory location initially when a logged reading is recalled...

Страница 4: ...ss to accept the selected setup option 5 Buttons NOTE Press all the dual function buttons momentarily to activate the Upper Row Function and press these buttons for more than 1 second to activate the...

Страница 5: ...display shows and LOG 2 Press to display the logging interval choices 3 Press or until the display shows the logging interval you want and then press to select 7 Setting the Date 1 Entering Setup will...

Страница 6: ...erature being measured is outside the thermocouple s valid range 9 Type Color Type Color J Black T Blue K Yellow E Purple 4 If you selected a user defined logging interval 1 second to 23 hours and 59...

Страница 7: ...set of initial conditions when logging continuously It stores 9 999 sets of temperature readings and initial conditions when logging individual points manually 11 Holding the Displayed Temperature 1...

Страница 8: ...Starting and Stopping Logging Memory clear and PC communications are in accessible during logging Recall function is enabled during logging 1 Set the time and the logging interval see How to Change Se...

Страница 9: ...ery 4 Reattach the battery compartment cover to the meter and reinstall the screws 15 These guidelines will help you attain long and reliable service from your meter 1 Calibrate your meter annually to...

Страница 10: ...cified for ambient temperatures between 64 F 18 C and 82 F 28 C for a period of 1 year The above specifications do not include thermocouple error 16 Specifications Environmental General Thermocouple O...

Отзывы: