CFK1 – OpTIONAL ChEMICAL FILTEr KIT (NOX FILTEr)

CO sensors may respond to gases other than CO, resulting in slightly elevated readings. On low NOx 
equipment this impact will be very small, often less than the tolerance of your meter. If you anticipate high 
levels of SO2 or NOx, UEi recommends using our CFK1, in-line chemical filter. The CFK1 will remove NOx and 
SO2 from the gas sample, allowing a true view of the CO only. The CFK1 is compatible with most portable CO 
and Flue Gas Analyzers, installs on the sample line before the instrument, and need for replacement is easy 
to determine.

AMBIENT AIr MEASurEMENTS

When testing for CO in rooms, make sure the meter has been switched on in outdoor fresh air before the 

meter is calibrated.

After the countdown is finished and the meter is correctly set up, take the meter into the room to be tested, 

make sure its pump is on and use the probe to draw in ambient air for the meter to measure.

•  Only switch on the meter in OUTDOOR, FRESH AIR
•  Do not exceed the meter’s internal temperature operating range
•  Do not put the meter on a hot surface
•  Do not expose yourself and others to dangerous levels of CO

prINTINg TEST INFOrMATION

An optional accessory for the CO95 is an infrared thermal printer. Read the manual supplied with the printer 

prior to operation. Connection to the CO95 is detailed below.

• 

Infrared thermal printer

 - this does not require

a cable to transmit the data but uses an infrared

(IR) link similar to a TV remote control. The

IR emitter is positioned on the top of the CO95

and the bottom of the printer. Ensure they are

pointing at each other and within 3 feet (1 meter),

with no obstructions in the way. Data may be

lost if transmission is interrupted. Keep the CO95

pointing at the printer until the printout has

finished.

During tests the CO95 will print data on request. With the 

analyzer showing the “

MAIN dISpLAy

” press 

“ 

 “ and current data will be sent to the printer.

The display will show “

prINTINg

” until data transmission 

is complete.

UEI CO95

503 644 8723

06-26-02

14:27

SMOKE:

NATURAL GAS

CO 

PPM 00

CO MAX 

PPM

TEMP. F  ----

Содержание CO95

Страница 1: ... O N M O N O X I D E D E T E C T O R CO95 INSTRUCTION MANUAL ENGLISH Carbon Monoxide Detector 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...easurements 6 CFK1 Optional Chemical Filter Kit NOx Filter 7 Ambient Air Measurements 7 Printing Test Information 7 After Using The Meter 8 Troubleshooting 8 Specifications 8 Maintenance 9 Periodic Service 9 Cleaning 9 Battery Replacement 9 Electromagnetic Compatibility EMC 10 Glossary of Terms 10 Warranty and Service Information 12 ...

Страница 3: ...s CO Max capture CO accurate to 5 of reading Temperature measurement Auto zero at power up Link to optional infrared printer Safety Notes Before using this meter read all safety information carefully In this manual the word WARNING is used to indicate conditions or actions that may pose physical hazards to the user The word CAUTION is used to indicate conditions or actions that may damage this ins...

Страница 4: ...r If the meter will not auto calibrate its sensors need to be replaced or recalibrated by an authorized repair center When countdown is finished the displays top line shows the last selected function and the bottom line displays whatever the rotary dial is turned to Press button again to switch the meter OFF The display counts down from 30 with the pump on to clean the sensors with fresh air If th...

Страница 5: ...T key Press HOLD ENTER again for live measurements 6 Menu Rotate the dial to MENU and use the UP or DOWN buttons to select the following function for change 1 Time Uses Military time as standard 7am 07 00 7pm 19 00 2 Date Select from MM DD YY YY MM DD or DD MM YY 3 The display s top line select from the settings listed below 4 Temperature units F or C 5 The display screen contrast 6 Language selec...

Страница 6: ...the CO95 s probe into the appliance s sampling point The ideal sampling point is at least two flue diameters downstream of any bend Put the probe tip in the flue center With balanced flue make sure the probe is positioned far enough into the flue so no air can back flush into the probe Make sure you do not exceed the meter s operating specifications In particular Do not exceed the probe s maximum ...

Страница 7: ... Only switch on the meter in OUTDOOR FRESH AIR Do not exceed the meter s internal temperature operating range Do not put the meter on a hot surface Do not expose yourself and others to dangerous levels of CO Printing Test Information An optional accessory for the CO95 is an infrared thermal printer Read the manual supplied with the printer prior to operation Connection to the CO95 is detailed belo...

Страница 8: ... not giving correct output Meter does not respond to flue gas Water trap drain plug not installed Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Erratic temperature readings or 000 on display Temperature plug reversed in socket Faulty connection or break in cable or plug Specifications General specifications Parameter Resolution Accuracy Range Tempera...

Страница 9: ...or weak battery from the meter and dispose of it in accordance with your local disposal regulations Batteries can leak chemicals that corrode electronic circuits If your meter is not going to be used for a month or more remove and store the battery in a place that will not allow leakage to damage other materials To install a new battery follow these procedures 1 Turn off the meter 2 Remove the rub...

Страница 10: ...erference has occurred and this advice is only given to satisfy the requirements of the Directive This product has been tested for compliance with the following generic standards and is certified to be compliant EN 50081 1 EN 50082 1 Specification EC EMC KI C100A details the specific test configuration performance and conditions of use Glossary of Terms TEMP Temperature measured by the probe in Fa...

Страница 11: ......

Страница 12: ...luding but not limited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A pur chase receipt or other proof of original purchase date will...

Отзывы: