GENERAL MAINTENANCE

PRE TEST CHECKLIST

SETTING INLET TEMPERATURE

ANALYZER CONNECTIONS

EMPTYING & CLEANING THE IN-LINE WATER TRAP

CHANGING THE PARTICLE FILTER

•  Re-certify your instrument annually to ensure it meets original performance specifications
•  Keep your instrument dry . If it gets wet, wipe dry immediately . Liquids can degrade electronic circuits
•  Whenever practical, keep the instrument away from dust and dirt that can cause premature wear
•  Although your instrument is built to withstand the rigors of daily use, it can be damaged by severe impacts . Use reasonable caution  

 

when using and storing the meter

•  Turn on and zero the analyzer without the flue probe connected to use ambient temperature
•  Connect flue probe thermocouple to T1 during zero countdown to store probe tip temperature as inlet (ducted system)
•  Connect a thermocouple to T2 to measure inlet during test .

The in-line water trap should be checked and emptied on a regular basis . Water vapor will condense in 

the probe line, which may cause the water trap to fill suddenly if the probe is moved . Care should be 

taken at all times .

Carefully remove the rubber plug from the bottom of the water-trap housing . Dispose of the condensate 

in a suitable drain, care must be taken as it could be acidic . If condensate spills onto the skin or clothing, 

clean off immediately using fresh water, seek medical advice if problems occur . Ensure plug is replaced 

before performing combustion tests . Note: O2 reading will be high if the Water Trap Plug is not in place .

This is a very important part of the analyzer and should be changed regularly . It prevents dust and dirt 

particles from entering the pump and sensors that will cause damage . The filter MUST be changed when 

it appears discolored on the inner surface .

Remove water-trap assembly from the analyzer as shown above . Remove the filter and plastic holder 

from the housing . Discard the filter element but keep the holder to fit to the new filter . Clean the inside 

of the filter housing with a suitable soft cloth . Fit the holder onto the new filter element and then insert 

into the housing . Refit the housing onto the analyzer .

•  Clean particle filter 
•  Water trap and probe line are empty of water
•  Power on and zero
•  All hose and thermocouple connections are properly secured

•  Flue gas probe is sampling ambient FRESH air
•  Water trap is fitted correctly to the instrument
•  Flue temperature plug is connected

WARNING!

Turning the pump off while the probe is in the flue will leave toxic gases inside 

the analyzer . Once data has been printed or copied it is advisable to purge the 

unit with fresh air as soon as possible . To do this remove the probe  from the flue 

and turn ON the pump . Always allow the readings to return to zero (20 .9 for O2) 

prior to shutting the unit off . The meter will not switch off until the CO reading is 

below 20 ppm .

WARNING!

The probe will be hot from flue gases . Remove the probe from the flue and allow 

it to cool naturally . Do not immerse the probe in water, as this will be drawn into 

the analyzer and damage the pump and sensors . Once the probe is removed from  

the flue and the readings have returned to ambient levels hold down “On/Off” 

and switch off the analyzer . The instrument will count down from 30 to switch 

off . If you pressed “On/Off“ by mistake, pressing “Send“ will return you to normal 

operation .

NOTE: Take care when inserting 

the temperature probes as the 

pins are polarized . Insert with the 

smaller pin (+) to the right . 

www.

.com

[email protected]

1.800.561.8187

Содержание C255

Страница 1: ...Enter I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS O2 9 7 COp 28 P 1 47 TFf 87 4 C255 C257 1 800 547 5740 Fax 503 643 6322 www ueitest com Email info ueitest com INSTRUCTION MANUAL ENGLISH Long Life Industrial EOS Analyzer www com information itm com 1 800 561 8187 ...

Страница 2: ...on Measurement Terms 14 Specifications 15 Electromagnetic Compatibility 15 Limited Warranty 16 DISPLAY SYMBOLS Excess Air Theoretical amount of air in excess of level needed to completely burn fuel Efficiency Calculated combustion efficiency based on net temperature oxygen and fuel selected Losses Losses calculated from oxygen and type of fuel Total NOx Calculated total Oxides of Nitrogen Ratio CO...

Страница 3: ...ssure SEND ENTER to Select and Print press and hold to log current values Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Battery Charge AC Adapter Flue Gas Inlet Connection Water Trap Drain 4 Line Backlit Display On Off Button Rotary Test Selector Dial Battery Compartments in back under boot AC Power Charge Indicator Light Temperature Connections Flue Probe ...

Страница 4: ...L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MENU SETUP C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MENU SETUP C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter SE...

Страница 5: ...MP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter LANGUAGE ENGLISH C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter SETUP NOx C225 L O N G L I F E E O S A N A LY Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter NOx 05 _ C225 L O N G L I F E E O S A N A L Y...

Страница 6: ...O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter RESET GAS ZERO IN FRESH AIR C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter FRESH AIR PURGE TIME 90 C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test T...

Страница 7: ...XCH Heat exchange integrity test TEMP and ROOM CO tests Press ENTER to view Press s UP or t DOWN to select VIEW DEL ALL or EXIT Select View to display logs Select Del All to erase records or Exit to move to the previous screen To delete records select Del All A prompt of Del All with YES shown prior to erasing records Press SEND ENTER to continue Press and hold s UP to change upper displayed param...

Страница 8: ...N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter AUX LINE 1 NOp_ N F C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MENU SETUP C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MENU SETUP C225 L...

Страница 9: ...ack adjust cone to aprox 2 from end of probe Rotate dial to Flue Test Rotate dial to Status The status menu shows battery life time date and the number of days until suggested re certification CAL Rotate dial to Aux The auxilliary menu shows user defined parameters To set up the auxilliary menu see instructions on page 8 Press SEND to print or press and hold SEND to log results Logged results are ...

Страница 10: ...p Room CO Up Down Enter I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS O2 9 7 COp 28 P 1 47 TFf 87 4 Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter C225 I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS CO2 6 2 NOp N F EFC 98 2 X 90 0 Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter C225 I N D U S T R I A L E O S A N A L Y Z E R CO2 ...

Страница 11: ...urning on and after readings have stabilized press the SEND button This will store the Pre Blower test segment After the blower turns on press PUMP to start the Post Blower test The meter will wait 60 seconds and then record the Post Blower values for CO O2 and Excess Air Test results will automatically be stored to exchange reports Report includes both Pre and Post Blower test segments and differ...

Страница 12: ... page 7 Press the PUMP button to start the test CO readings peak will be logged every 2 minutes for a 30 minute time span To see logged results see page 7 Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter C225 I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS TEMP T1f 000 T2f 000 Tf 000 C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Te...

Страница 13: ...DE after a 4 second countdown and automatically zero the pressure sensor Analyzer model number time date and countdown will show C225 L O N G L I F E E O S A N A LY Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MANOMODE Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter C225 I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS P ...

Страница 14: ...n O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Supply and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit Switch Pressure Switch Verify proper operation Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils 90 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Effici...

Страница 15: ...o 9 O2 Stack Temperature 325 to 500 F Draft Water Column Inches 02 to 04wc Carbon Monoxide parts per million 100ppm Gas Fired Power Burners Oxygen 3 to 6 O2 Stack Temperature 275 to 500 F Stack Draft Water Column Inches 02 to 04wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Oil Fired Burners 2 Oil Fuel Oxygen 4 to 7 O2 Stack Temperature 325 to 600 F Stack Draft...

Страница 16: ...g the pump and sensors that will cause damage The filter MUST be changed when it appears discolored on the inner surface Remove water trap assembly from the analyzer as shown above Remove the filter and plastic holder from the housing Discard the filter element but keep the holder to fit to the new filter Clean the inside of the filter housing with a suitable soft cloth Fit the holder onto the new...

Страница 17: ... local disposal regulations WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrument This could alter the protection from personal injury this meter provides to the operator Perform only those maintenance tasks that you are qualified to do Remove the probe from the flue and allow t...

Страница 18: ...and the stack tempera ture and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate the efficiency as adjustments are being made Combine this with a printout and you are able to provide a before and after comparison of the combustion equipment in relatively little time ...

Страница 19: ...F C 0 3 reading 3 6 F 2 0 C Inlet Temperature Internal sensor 32 122 F 0 50 C 1 0 F C 0 3 reading 1 8 F 1 0 C Inlet Temperature External sensor 32 1112 F 0 600 C 1 0 F C 0 3 reading 3 6 F 2 0 C Gas Measurements Carbon Monoxide CO 1 0 60 ppm 1ppm 3ppm 61 4 000ppm nom 1ppm 5 of reading 4 001ppm to 10 000ppm 1ppm 10 of reading 20 000ppm max for 15 min 1ppm Not specified Carbon Dioxide CO2 1 0 20 0 1 ...

Страница 20: ... products All recertification services are warranted for one year Save time with online recertification check in by visiting www ueitest com and click the SERVICE REQUEST banner and filling out the form Our current recertification prices can be found on the SERVICE section of our web site Up front pricing means NO surprises LIMITED WARRANTY The C255 combustion analyzer is warrantied for five years...

Отзывы: