32
MENU
REMOVE PROBE FROM EXHAUST
PRESS ENTER TO ZERO CAL
12.52.30 pm
03/22/15
Ensure the probe is removed from the vehicle exhaust and is sampling fresh air, in the garage environment.
This should be 2 meters above the ground. Once this is done press
to activate the pump.
WARNING :
The sensors will only be reset if the probe is sampling fresh air for at least 60 seconds.
Once the zero is complete the screen will return to the
MAIN DISPLAY
.
9. PRODUCT SPECIFICATION FOR HANDHELD AUTOPLUS 4-2 & AUTOPLUS 5-2
Parameter
Resolution
Accuracy
Range
Carbon Monoxide
(Infrared)
0.01 %
± 5 % of reading *1
± 0.06 % volume *1
0-10 %
Over-range 20 %
Oxygen
(fuel cell)
0.01 %
± 5 % of reading*1
± 0.1 % volume *1
0-21 %
Over-range 25 %
Hydrocarbon
(Infrared)
1 ppm
± 5 % of reading *1
± 12 ppm volume *1
0-5000 ppm
Over-range: 10,000 ppm
Carbon Dioxide
(Infrared)
0.1 %
± 5 % of reading *1
± 0.5 % volume *1
0-16 %
Over-range: 25%
Nitric Oxide*²
(fuel cell)
1 ppm
± 5% of reading *1
± 25ppm of volume*1
0-1500ppm
Over-range: 5000 ppm
Carbon Monoxide
Corrected COK
0.01%
Calculated
0-15%
Lambda
AFR (Petrol)
(LPG)
0.001
00.01
0.8 – 1.2
11.76 – 17.64
12.48 – 18.72