GB-50
Malfunction
Possible reasons
Corrective action
Display
Er- 300
Excessive cuff pressure
The cuff inflated to the maximum by
movement of the arm or body
Repeat measurement, no moving the
arm or speaking
Display
Err -1
Measurement error
Measurement could not be carried out
due to movement of arm or body
Repeat measurement, no moving the
arm or speaking
Display
Err -2
Inflation error
Pressure built up too quickly or too
slowly
Check position of air plug, repeat
measurement
Error was recognized
Repeat measurement, no moving the
arm or speaking
Display
0
0
Defective measurement
Start/stop button accidentally pressed
while changing battery
Reboot device with Start/stop button
Measurement interrupted and
cuff deflated and re-inflated
Device detects defect or diastolic value
too low
Device shows no defect
Repeat measurement
Movement during measurement
Interrupt measurement if necessary
New measurement after 5 minutes
break
Appendix
G
english
Содержание visomat comfort 20/40
Страница 6: ...deutsch A D 6 Gerätebeschreibung ...
Страница 29: ...D 29 Anhang G deutsch ...
Страница 34: ...A GB 34 Display unit description english ...
Страница 60: ...français A F 60 Descriptif de l appareil Indications de contrôle ...
Страница 70: ...français F 70 D Annexe ...
Страница 74: ...italiano A I 74 Descrizione dello strumento indicazioni per il controllo ...
Страница 84: ...italiano I 84 Appendice D ...
Страница 88: ...A E 88 Descripción del aparato español ...
Страница 98: ...español E 98 D Apéndice ...
Страница 102: ...ελληνικά A GR 102 Περιγραφή της συσκευής ...