1. Error messages
Error
Possible cause
Remedy
Display Err - 300
Excessive cuff
pressure. The cuff
was inflated to its
maximum due to the
movement of the
arm or body.
The air hose may be
kinked or the air plug
blocked.
- Repeat the measurement
- Do not move your arm
- Do not speak
Inspect the hose line and air
plug for blockages
Display Err - 1
Measurement error;
measurement could
not be carried out
due to movement or
speaking.
- Repeat the measurement
- Do not move your arm
- Do not speak
The acquired values
lie outside the mea-
suring range of the
device (see „Technical
data“ on page 58).
Repeat the measurement
Display Err - 2
Inflation error;
pressure built up too
quickly or too slowly.
Check that the air connector
is fitted properly, repeat
measurement.
Fault has been iden-
tified. The cuff is not
correctly attached
- Repeat the measurement
- Do not move your arm
- Do not speak
Display Err - 3
Deflation rate too
high or too low.
Call the Service hotline if
this occurs repeatedly.
Display 0
0
Start/Stop button
was accidentally
pressed while the
batteries were being
replaced or while the
mains adapter was
being plugged in.
Switch unit off and then on
again using the Start/Stop
button.
Error
Possible cause
Remedy
Arrhythmia, mo-
vement disorders,
trembling, wobbling
(objects), deep bre-
athing etc.
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device‘s
pulse beep signals. Possible
influences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed with
your doctor!
The motion control
detected movement
during the measure-
ment.
- Repeat the measurement
- Do not move your arm
- Do not speak
Measured values/
pulse values are
extremely high or
low.
Too little rest before
taking the measure-
ment. Movement or
talking during the
measurement.
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 43.
Incorrect measure-
ment posture
Repeat the measurement
and maintain the correct
posture. (Page 47)
Measurements
taken in direct
succession
produce different
measured values.
Blood pressure is not
a fixed value. It can
go up or down by
more than 20 mmHg
within a few minutes.
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Measurement was
interrupted, the
cuff deflated to 0
and the measure-
ment restarted.
Device detects a fault
or that the diastolic
value is too low.
Device shows no fault.
Repeat measurement
Movement during
measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes‘ rest.
No display after
device is switched
on
Batteries inserted
incorrectly
Check position of batteries.
Batteries flat
Replace batteries.
Battery contacts
dirty.
Clean battery contacts with
a dry cloth.
Dashes in display
after pressing
Memory button
No measured values
have been stored.
Repeat measurement