30
RMT
Align the pins located at the rear of the module with the four jacks in the front end of the
Metered Pistol Housing (MPH) and plug in. Place the magnetic transducer on the test location.
TO CHARGE THE UP15,000
Plug recharger cable into recharger jack on the UP10000 and then plug the recharger
into a wall receptacle.
Make sure that the LED on the charger is blinking when recharging.
The LED remains solid when the battery is charged. The instrument may stay
connected to the charger without damaging the battery. Charge time is
approximately 4 hours.
WARNING:
Use the supplied UE Systems recharger (BCH-10L) only. Use of unauthorized
rechargers will void the warranty and may degrade or damage the battery.
WARBLE TONE GENERATOR / SENSITIVITY VALIDATION UNIT
(UE-WTG-1)
The Tone Generator has two functions.
Sensitivity Validation:
This is a procedure that should be incorporated BEFORE
performing any test with the Ultraprobe. It provides assurance that the instrument is
working properly to provide repeatable, reliable results for any inspections. This is a
procedure that is recommended for any Predictive Maintenance instrument used. If
performed regularly it will promote accuracy and test reliability throughout the PdM
program.
For detailed instructions, refer to Appendix A: “Sensitivity Validation Procedure”
As a method for locating large leaks when it is difficult to produce pressure
or vacuum.
To perform this test:
a.
Turn Tone Generator on by selecting either "LOW" for a low amplitude signal or
"HIGH" for high amplitude. When the Tone Generator is on, a red light (located
below the recharge jack in the front) flickers.
b.
Place the Warble Tone Generator within the test item/container and seal or close
it. Then scan the suspect areas with the TrisonicTM Scanning Module in the
Ultraprobe and listen for where the "warble" ultrasound penetrates.