TRIGGER SWITCH.

 This is located on the underside of the 

Ultraprobe 100.  The Ultraprobe 

is always “off” until the trigger switch is 

pressed.  To operate, simply press the trigger and hold; to turn the 
instrument off, release the trigger. 

 

SCANNING MODULE 

This module is utilized to receive air-borne ultrasound such as the ultrasounds 
emitted by pressure leaks and electrical discharges. To use, make sure it is 
plugged in to the front end of the metered pistol housing by aligning the plug 
with the receptacle and inserting it firmly.  It has approximately a 60° field of 
reception. 

 

TO USE THE SCANNING MODULE 

 

Plug in to front end 

 

Start with the sensitivity selection dial at maximum (0) 

 

Start to scan the test area

 

 
The method of air borne detection is to go 

from the “gross to the fine”. If there is too much ultrasound 

in the area, reduce the sensitivity, place the RUBBER FOCUSING PROBE (described below) over the 

scanning module and proceed to follow the test sound to its’ loudest point. If it is difficult to 

locate the 

sound due to a high intensity signal, keep reducing the sensitivity and following the meter to the loudest 
point. 

 
CONTACT (STETHOSCOPE) MODULE 

This is the module with the metal rod.  This rod is utilized as a “wave guide” that

 

 is sensitive to ultrasound generated internally such as with a pipe, bearing housing, steam 
trap or wall.  Once stimulated by ultrasound, it transfers the signal to a piezoelectric 
transducer located directly in the module housing.

 

TO USE THE STETHOSCOPE MODULE 

 

Align the pin located at the rear of the module with the jack in the front end of the 
Metered Pistol Housing and plug in firmly. 

 

Touch test area. 

 

As with the scanning module, go from the “gross” to the “fine.”  Start a maximum sensitivity on the 

Sensitivity Selection Dial and proceed to reduce the sensitivity until a satisfactory sound and meter 
level are achieved.

 

Содержание Ultraprobe 100

Страница 1: ...1 UE SYSTEMS INC THE ULTRAPROBE 100 MANUAL...

Страница 2: ...OVERVIEW COMPONENTS Introduction Basic Components of Your Kit Pistol Housing Scanning Module Contact Stethoscope Module Rubber Focusing Probe Long Range Module Headset WTG 1 Warble Tone Generator 2 2...

Страница 3: ...inspection through advance ultrasonic technology Before you begin testing it is advisable to familiarize yourself withthe basic components of your kit BASIC COMPONENTS OF YOUR KIT Contact Extension Ki...

Страница 4: ...er on and then stay off This is normal and has no relation to battery condition SENSITIVITY SELECTION DIAL There are eight 8 sensitivity levels which read out in related decibels of 0 to 70 As the dia...

Страница 5: ...uce the sensitivity place the RUBBER FOCUSING PROBE described below over the scanning module and proceed to follow the test sound to its loudest point If it is difficult to locate the sound due to a h...

Страница 6: ...s remove the module BEFORE attaching and removing the Long Range Module HEADSET This heavy duty headset is designed to block out intense sounds often found in industrial environments so that the user...

Страница 7: ...headphones A smooth continuous warbling sound should be heard If a beeping is heard instead then a full recharge of the Warble Tone Generator is indicated TO CHARGE THE WARBLE TONE GENERATOR Use the...

Страница 8: ...tempting repairs Be careful with loose hanging parts such as the wrist strap or headphone cord when inspecting near moving mechanical devices since they may get caught Don t touch moving parts with th...

Отзывы: