10,000SD

VALVE INSPECTION

Ultraprobe 10,000SD 

-

Valve Inspection: 

How To Guide

NOTE: 

UE DMS Software available for download at 

www.uesystems.com

SAFETY: WRIST STRAP WORN ALL TIMES

A, B, C, D, AND SOMETIMES E METHOD

The 

A,B,C,D and Sometimes E Method 

Contains 

Four 

(4) 

Easy Rules

For Valve Condition Diagnosis:

1. If Test Point 

C

is the Loudest, 

LEAK-BY

is Present.

2. If Decibel Lowers From Test Points 

-

D

, the Valve is 

CLOSED

.

3. If Readings are the 

Same Or Similar

, the Valve is 

OPEN

.

4. If 

D

is Higher Than A,B, And C, the Ultrasonic 

SOURCE IS DOWNSTREAM

. Incorporate Test Point 

E

(Check Downstream)

A

B

C

D

Содержание ULTRAPROBE 10,000SD

Страница 1: ...VALVE INSPECTION VISUAL INSTRUCTION ULTRAPROBE 10 000SD ...

Страница 2: ...Time Route Based OR Trouble Shooting Tactic in Assessing Valve Operation Being a Critical Tool for Reliability Centered Maintenance The Ultraprobe 10 000SD Allows for Analysis of Valve Operation Sustaining Mechanically Based Systems is One of the Most Expensive Utilities Assessing Real Time Deviations and Inefficiency Factors is an Advantage in Directly Understanding Root Cause in Order to Maintai...

Страница 3: ...s com SAFETY WRIST STRAP WORN ALL TIMES INITIAL SETUP To Power On the Ultraprobe 10 000SD Pull and Hold the Instrument Trigger Keep Trigger Pulled for Instrument to Remain ON The Unit Will Power On in the Main Display The Ultraprobe 10 000SD Powers On in Default Mode of 40KHz with the Sensitivity at Maximum 70 Unless Previously Changed ...

Страница 4: ...e 10 000SD for Valve Inspection Center the Frequency to 25KHz Using the Sensitivity Dial Click Dial Until KHZ is Flashing Then Spin to the Desired Frequency Once 25KHz is Reached Click the Sensitivity Dial Again to Set That Frequency Then Insert the Stethoscope Module STM or the Remote Magnetic Transducer RMT into the Module Port Modules are Removed By Pulling Module Straight Out of Port Without T...

Страница 5: ...m the Setup Menu Located in the Function Bar Press the Sensitivity Dial Until the Setup Menu is Flashing Then Press the ENTER Button Spin the Dial to find Menu 3 Application Select To Enter the Menu Click the Dial Again From Menu 3 Application Select Spin the Sensitivity Dial to Navigate to VALVES Then Click the Dial Again to Save That Selection Once VALVES is ENABLED Press the ENTER Button to Ret...

Страница 6: ...This Will Allow Inspectors to Create a Route s for Each Valve Within the System Each Route Can House Up to 400 Valves and Can Be Uploaded to the Ultraprobe 10 000SD For Organized Data Collection and Review Each Route is Stored Within the Plant Created Within the DMS Software and is Available for Upload Download and Review at Anytime NOTE the DMS Software is NOT Required if Using the Ultraprobe 10 ...

Страница 7: ...lect Data Make Contact With the First Test Point Using the A B C D And Sometimes E Method This Method Incorporates Four Individual Data Points Two 2 Upstream and Two 2 Downstream Per Valve to Uncover the Source of the Ultrasonic Signal For Each Test Point Adjust and Optimize the Sensitivity to Capture a Decibel dB Value and if Necessary Tune Frequency to Obtain a Good Quality Signal Always Ensure ...

Страница 8: ... the Condition of the Valve This Method is Utilized to Confirm the Source of Detected Turbulence This Simplistic Method Allows for Consistent Repeatable Testing 1 Test Point A is Three Times 3x the Diameter of the Piping Away From the Valve On the Upstream Side 2 Test Point B is One Times 1x the Diameter of the Piping Away From the Valve On the Upstream Side 3 Test Points C and D Mirror Test Point...

Страница 9: ...TIMES E METHOD The A B C D and Sometimes E Method Contains Four 4 Easy Rules For Valve Condition Diagnosis 1 If Test Point C is the Loudest LEAK BY is Present 2 If Decibel Lowers From Test Points A D the Valve is CLOSED 3 If Readings are the Same Or Similar the Valve is OPEN 4 If D is Higher Than A B And C the Ultrasonic SOURCE IS DOWNSTREAM Incorporate Test Point E Check Downstream A B C D ...

Страница 10: ... is a Critical Component for Organized Data Acquisition and Diagnoses When VALVE Application is Selected Enter the Setup Menu From the Function Bar Via the ENTER Button Enter Menu 2 Display Screens By Pressing the Sensitivity Dial Spin the Sensitivity Dial to ABCD Then Press the Dial Again to Save that Display Option The Main Display Will Now Show the Four 4 Necessary Test Points Marked A B C and ...

Страница 11: ...Spin the Dial A RIGHT Arrow Indicates the Sensitivity is Too Low the Dial Must Be Turned to the Right to Obtain a Decibel A LEFT Arrow Indicates the Sensitivity is Too High When the Sensitivity is Optimized a Decibel dB Will Populate On the Display Ready to Be Applied to That Test Point The Ultraprobe 10 000SD Display Indicates Which Test Point Value is Being Acquired Simply Press ENTER to Store T...

Страница 12: ... SOFTWARE For Program Support Download the DMS Software from UESystems com Once Downloaded Create Routes for Each Valve Within the System Designating Specific Inspection Processes and Methodology Directly Supporting Organization and Repair The DMS Software Allows for Data Storage and Report Generation From the Reports Tab Select the Desired Report Then Select Generate Selected Report ...

Отзывы: