VALVE INSPECTION
Ultraprobe 10,000SD
-
Valve Inspection:
How To Guide
NOTE:
UE DMS Software available for download at
www.uesystems.com
SAFETY: WRIST STRAP WORN ALL TIMES
INITIAL SETUP
To
Set Up
the Ultraprobe 10,000SD for
Valve
Inspection
, Center the Frequency to
25KHz
Using the Sensitivity Dial. “Click” Dial Until
“KHZ” is Flashing, Then Spin to the Desired
Frequency. Once
25KHz
is Reached,
Click
the
Sensitivity Dial
Again to
Set
That Frequency.
Then
Insert
the
Stethoscope Module
(STM) or
the
Remote Magnetic Transducer
(RMT) into
the Module Port. Modules are Removed By
Pulling Module
Straight Out
of Port Without
Twisting.
25 KHZ