PAGE 3
TO CLEAN THE FERRULE:
•
Unplug the test jumper
•
Unscrew the adapter until it spins freely, then pull
•
Inspect the ferrule with an approved microscope,
clean if necessary, and replace adapter
The LSSM utilizes a physical fiber connection at the output port. It ensures a steady power level for performing
insertion loss tests.
Ensure that any test jumpers are inspected and cleaned before plugging into the LSSM unit. If soiled or damaged
connectors are inserted, they can cause damage to the LSSM output port and the unit may need to be repaired.
OUTPUT PORT MAINTENANCE
LSSM
When the LSSM’s 2kHz modulation is active, the selected wavelength indicator will blink. The modulation
is recognized by optical power meters further down the fiber and indicated by a 2kHz notification on the
screen and a loud beep. The modulated tone can also be recognized through the fiber jacket when a Fiber
Identifier is used.
USING THE 2KHZ FUNCTION
Test Jumper has physical contact with the
ferrule inside the LSSM
The Fiber Identifier introduces a bend into the fiber which leaks light from the core onto the detector of the
power meter. When using a Fiber Identifier, expect a 30 to 35dB offset in core power.
Содержание TEK-6
Страница 2: ...SECTION ONE LSSM LASER LIGHT SOURCE QUICK START GUIDE NORTH AMERICA...
Страница 9: ...SECTION TWO LSSM LASER LIGHT SOURCE TECHNICAL GUIDE NORTH AMERICA...
Страница 12: ...SECTION THREE OPM1 OPTICAL POWER METER QUICK START GUIDE NORTH AMERICA...
Страница 19: ...SECTION FOUR OPM1 OPTICAL POWER METER REFERENCE TEST GUIDE NORTH AMERICA...
Страница 31: ...SECTION FIVE HFM1 HANDHELD INSPECTION MICROSCOPE QUICK START GUIDE NORTH AMERICA...
Страница 34: ...SECTION SIX VFF1 VISUAL FAULT FINDER QUICK START GUIDE NORTH AMERICA...