PAGE 3 OF 9
0621 IH-8109
LOW BATTERY INDICATION
The display will show
Lo
when the batteries are weak
and need to be replaced.
AUTO SHUT-OFF
To extend battery life, the scale will automatically turn off
after approximately four minutes if no active weighing is
occurring. This feature is active with battery operation only.
CALIBRATION
For best results, calibrate the scale at regular intervals,
especially if using for prolonged periods. Calibration
weights are not provided with the scale.
The message
E
will appear
if the calibration steps are
not followed or the wrong
weight was used.
1. Press and hold the
g
button until
7ENU
is
displayed. Release the
button, first menu
CAL
will be displayed.
NOTE: The calibration process can be aborted
by pressing
g
2. Press
0
to start span calibration.
3. The display shows
0
.
With no weight on the
pan, press the
0
button to establish the
zero point. The display
shows
--C--
while the
scale stores the zero
load signal.
4. The display shows the span calibration point, for
example, 200g. Place the specified calibration
weight on the pan and press the
0
button. The
display shows
--C--
while the scale stores the
calibration point signal.
5. The display shows
[-dONE-]
and returns
to the normal weighing
mode.
BACKLIGHT
The backlight settings can
be adjusted using the
menu. Press and hold the
g
button until
7ENU
is displayed. Release the
button, first menu
CAL
will
be displayed. Press
g
to
select desired option.
MENU OPTION
SETTING
b.Auto
Scale turns on light auto-
matically with any
button or weighing opera-
tion.
b.ON
Backlight is always on.
b.OFF
Backlight is always off.
Once desired setting is displayed, press
0
to select
and return to current weighing mode.
OPERATION CONTINUED
CALIBRATION VALUES
MODEL #
CALIBRATION POINT (g.)
H-8109
200
H-8110
1,000
H-8111
2,000