PAGE 5 OF 6 

1114 IH-4729

SEALING

PHYSICAL SEALS

For jurisdictions that use the physical sealing method, 

the local weights and measures official or authorized 

service agent must apply a security seal to prevent 

tampering with the settings. Refer to the illustrations 

below for sealing methods. (See Figures 2-3)

AUDIT TRAIL SEAL

For jurisdictions that use the audit trail sealing method, 

the local weights and measures official or authorized 

service agent must record the current configuration and 

calibration event counter values at the time of sealing. 

These values will be compared to values found during a 

future inspection.

NOTE: A change to an event counter value is 

equivalent to breaking a physical seal.

The audit trail uses two event counters to record 

changes to configuration and calibration settings:

•  The configuration event counter (CFG) will index 

by 1 when exiting the menu if one or more of the 

following settings are changed: Legal for Trade, 

Calibration Unit, Capacity, Graduation, Power On 

Unit, Zero Range, Auto Zero Tracking, Expanded 

Mode, Count Mode, Kilogram Unit, Pound Unit, Gram 

Unit, Ounce Unit, Pound Ounce Unit, Stable Only. 

Note that the counter only indexes once, even if 

several settings are changed. The configuration 

event counter values range from CFG000 to CFG999. 

When the value reaches CFG999, the count starts 

over at CFG000.

•  The calibration event counter (CAL) will index by 1 

when exiting the menu if a Span Calibration, Linearity 

Calibration or GEO setting change is made. Note 

that the counter only indexes once, even if several 

settings are changed. The calibration event counter 

values range from CAL000 to CAL999. When the value 

reaches CAL999, the count starts over at CAL000.

The event counters can be 

viewed by pressing and 

holding the 

MENU

 button. 

While the button is held, 

the display will show MENU 

followed by AUd It.

1.  Release the button when 

AUd It is displayed to view 

the audit trail information.

2.  The audit trail information 

is displayed in the format 

CFGxxx and CALxxx.

3.  Then the indicator returns 

to normal operation.

OPERATION CONTINUED

Wire Seal

Paper Seal

Figure 2

Figure 3

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

CALIBRATION

MODE

MAINTENANCE

CAUTION! Disconnect the unit from  

the power supply before cleaning.

CLEANING

•  The housing may be cleaned with a cloth 

dampened with a mild detergent if necessary.

•  Do not use solvents, chemicals, alcohol, ammonia 

or abrasives to clean the housing or control panel.