PAGE 3 OF 3
0616 IH-2780
OPERATION
TROUBLESHOOTING
BUTTON FUNCTION
Press the ORIGIN button to zero the micrometer.
CAUTION! The LCD automatically turns off if it
has been idle for 20 minutes. To turn on the
LCD, turn the thimble.
ZEROING THE MICROMETER
IMPORTANT! When the battery has been
installed, first press the
ORIGIN
button, but
do not rotate the thimble while the preset
values are being set. (See Figure 1) Rotating
the thimble during this time may result in the
failure of setting the default settings. Reinstall
the battery if you moved the thimble during
this time.
• Before measuring check the origin (zero point)
according to Figure 3.
• Remove dust/oil from the measuring faces before
zeroing the micrometer.
• It is recommended that the micrometer is held in the
same position and under the same conditions for
both zeroing and actual measurement.
Zero the micrometer by following the procedure below:
(See Figure 3)
1. Turn the thimble to gently bring the spindle into
contact with the anvil. Apply the rated measuring
force using the ratchet stop/friction thimble.
2. Press the ORIGIN button.
3. If 0.000 is displayed and "P" is flashing, press the
ORIGIN button again. The display value will not
change while "P" is flashing.
4. "P" disappears, indicating the micrometer has zeroed.
CAUTION! If the
ORIGIN
button is accidentally
pressed during measurement, press the
ZERO/ABS
button to restore. Zero the micrometer again if
measurement function is still not restored.
ERROR MESSAGE
OPERATING ISSUE
RECOMMENDATIONS
The battery is depleted.
Replace the battery.
Err-oS
Error indicates when a counting error
occured due to noise or moving
spindle too fast.
Reset the battery and zero the
micrometer.
Err-S
Error indicates when a counting
error occured due to an error in the
electronic unit during the initial setting
of the micrometer.
Reset the battery and zero the
micrometer.