PAGE 4 OF 8
0313 IH-3827
OPERATION CONTINUED
CHANGING UNITS OF MEASURE CONTINUED
To select a unit of measure, press and hold the
Unit button, then release it when the desired unit is
displayed.
NOTE: If the desired unit is not displayed, it
must be turned on in the Unit menu.
CHANGING APPLICATION MODES
The Adventurer Pro can be configured to operate in
various application modes. The Mode Sub-menu is used
to enable or disable a specific application mode.
Press and hold the Mode button and release it when the
desired mode is shown on the secondary display.
NOTE: If the desired mode is not shown, it must
be selected in the mode menu. Press and hold
the Mode button until it appears on the display.
PRINTING DATA
Press the Print button to send the displayed value to the
COM port. Note: The port must have Auto Print set to OFF.
CHANGING SETTINGS
To change a menu setting, perform the following steps:
ENTER THE MENU
Press and hold the Menu button until MENU appears
on the primary (7-Segment) display. Release the button
and the first sub-menu appears on the secondary
(14-Segment) display (blinking).
NOTE: When the secondary display is blinking,
press Yes to accept. Press No or Back or view
another selection.
SELECT THE SUB-MENU
Press the No button to select the next sub-menu or
the Back button to select the previous sub-menu in
the loop. Press the Yes button to select the sub-menu
that appears on the secondary display (blinking). The
sub-menu name now appears on the primary display
and the first menu item in the sub-menu now appears
on the secondary display (blinking).
SELECT THE MENU ITEM
Press the No button to select the next setting or the
Back button to select the previous setting in the loop.
Press the Yes button to select and save the setting on
the secondary display (blinking) as the new menu item
setting.
EXIT THE MENU
Press the Exit button at any time to quickly exit the menu
or select END sub-menu and press the Yes button.
APPLICATION MODES
WEIGH
1. Press and hold Mode until
the display shows WEIGH,
then release the button.
2. Press Zero to zero the
balance.
3. Place objects to be weighed on the pan to display
the weight. The example displays a 200 gram weight.
4. Press and hold Mode until
Count is displayed, then
release the button.
PARTS COUNTING
Use the Parts Counting Mode to count samples of
uniform weight.
Press and hold Mode until Count is displayed, then
release the button.
ESTABLISH AN AVERAGE PIECE WEIGHT (APW)
1. Each time a new type
of part is counted, the
nominal weight of
one piece (Average
Piece Weight) must be
established using a small quantity of pieces. This
APW is stored until replaced by another APW.
2. If an APW has already been established, Count is
displayed with Clear APW? blinking.
3. Press No to use the previously saved APW.
4. Otherwise, press Yes to establish a new APW.
5. The current weight (0.00)
is shown on the primary
display and the default
sample size (PLACE 10) is
shown on the secondary display.
(Clear APW? blinking)