![Tyco kENDALL KANGAROO ePump Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010033.webp)
KANGAROO
e
Pump
™
Enteral Feeding Pump
figure . feed error screen.
flush error
(Adjustable Audible Volume)
The
flUsH eRRoR
screen appears when the flushing solution is no longer being delivered because the flush bag is empty or
because of a clog between the pump and flush bag. Check the flush line to find the occlusion causing the blockage. If the occlusion
cannot be cleared, load a new Pump Set, prime it, and restart the feeding, Figure 29
.
If the error still cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service.
figure . flush error screen.
flow error
(Adjustable Audible Volume)
The
floW eRRoR
screen appears when either the feeding or the flushing solution is no longer being delivered because of a clog
between the pump and the patient. Check the line to find the occlusion causing the blockage. If the error cannot be resolved, check
the valve pocket in the Pump Set loading area for dampness or dirt, which may be a cause for a false error. Clean and dry the valve
pocket. If the error still cannot be resolved, load a new Pump Set, prime it, and restart the feeding, Figure 30.
If the error still cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service.
figure 0. flow error screen.
Pump set Dislodged error
(Fixed Audible Volume)
The
PUMP seT DIsloDGeD
screen will appear if the black ring retainer (MISTIC) is not properly loaded in the MISTIC pocket in the
Pump Set loading area. Check and correct the positioning of the MISTIC retainer if possible. If the error cannot be resolved, load a
new Pump Set, prime it, and restart the feeding, Figure 31.
If the error cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service
.