![Tyco kENDALL KANGAROO ePump Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010028.webp)
KANGAROO
e
Pump
™
Enteral Feeding Pump
section VII — Cleaning
Cleaning should be performed as needed. It may also be desirable to define cleaning intervals based on knowledge of the
environment in which the pump is used. Only personnel trained in the cleaning of medical devices should perform cleaning.
CaUTIon: Do not immerse pump or power cord in water or other cleaning solution; clean using a damp (not wet) cloth
or sponge. failure to follow the cleaning procedures described herein could result in hazards to users. as with any
aC powered electrical device, care must be taken to prevent liquid from entering the pump to avoid electrical shock
hazard, fire hazard, or damage to electrical components.
If any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel trained in
servicing KANGAROO ePump enteral feeding pumps:
• wetting of the pump’s power cord or leakage into the pump interior during cleaning.
• spillage of large amounts of formula onto the pump exterior, or any spillage onto the power cord.
General Cleaning Directions
Cleaning of KANGAROO ePump enteral feeding pumps must be performed as follows:
CaUTIon: Disconnect pump from aC power source before cleaning. after cleaning, do not connect to aC power source
until pump and power cord are thoroughly dry.
• A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and hypochlorite
mixture, however, repeated cleaning with this solution can damage the plastic pump case. Isopropyl alcohol, applied with a damp
(not wet) cotton swab, may be used for cleaning difficult-to-reach areas, however, it should be used sparingly because repeated
cleaning may damage the case.
• DO NOT USE strong cleaners such as Spray-Nine™*, Phisohex™*, Hibiclens™*, or Vesta-Syde™* because damage to the pump case
housing can result.
Directions for Cleaning Pump Housing
• Refer to General Cleaning Directions before starting.
• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as possible.
• Avoid excess moisture near pole clamp area.
• Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.
Directions for Cleaning Pump Power Cord
• Refer to General Cleaning Directions before starting.
• Unless soiling is observed, the power cord should not be cleaned.
• If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with a cloth
dampened with isopropyl alcohol.