14.5 CABLIBRATION S/W INSTALLATION
If you are connec
ti
ng an X-Rite, CA-210, CS-200, or an Eye-One, connect the probe with the
PC through the USB port.
Those probes which connect with USB should install the device driver
off
ered by the
measurement device company.
(The XVM-245W supports the Klein K-10, Konica-Minolta CA-210, CS-200 and X-Rite i1
Display3 probes)
Put the measurement device facing toward the center of the monitor.
Set the input source of the monitor as SDI-A.
FCM Series, LHM Series: A
ft
er input the DVI input signal, set the input source of the monitor
as DVI
If you pre-heat (or turn on) the monitor for more than 30 minutes before the calibra
tion
,
you can get a more accurate result.