Page 74
7.
Disassemble or sufficiently loosen multiple-part instruments prior to
packaging to permit the sterilizing agent to come into contact with all
parts of the instrument.
8.
Tilt on edge items prone to entrap air and moisture, e.g. hollowware, so
that only minimal resistance to removal of air, the passage of steam
and condensate will be met.
9.
Once a week, use a biological spore test indicator in any load to make
sure sterilization is performed.
10.
Make sure that all instruments remain apart during the sterilization
cycle.
11.
Load the basket loosely to capacity.
9.2
Liquids
1.
Use only heat- proof glass containers, filled to 2/3 capacity.
2.
For Slow exhaust cooling (without air) the glass container should be
covered but unsealed.
3.
Place the two temperature sensors into two separate liquid containers.
These are used to control the program temperature and ensure the
safety of the operating cycle.
9.3
Loading
The loading of goods and instruments is done by means of two stacked
baskets. The baskets are provided with handles for the convenience of
the operator.
There are 2 types of baskets:
1.
Baskets that are fully perforated.
2.
Baskets that are not perforated except one row of holes adjacent to the
basket’s top. These baskets are intended for waste cycles, to avoid
clogging of the vessel’s drainage pipe by overflowing liquids.