Page 45
needed to prevent swelling or warping of soft plastic items
.
Chamber
fan is used to mix air with steam.
•
Sterilization:
Sterilization is controlled in a way that the sterilization
pressure remains approx. 30% above the theoretical pressure
corresponding to the same temperature according to the steam table.
•
Cooling:
Forced cooling to the required end temperature, chamber
pressure remains at approx. 300 kPa and the cooling valve is opened.
•
Fast Exhaust:
The steam is exhausted out of the chamber at a fast
rate until pressure decreases to ambient pressure.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.17
Program 17: Glass Test
This cycle is intended to check the durability of the bottles for liquids
when the manufacturer recommends autoclaving at temperatures of
121ºC (250
F) for 20 minutes.
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F).
•
Sterilization time:
20
minutes.
Operations sequence:
•
Heating:
Steam enters the chamber and heats it up until the
sterilization temperature is reached.
During the interval of the heating stage, between 100°C and
121°C, the heating rate is kept to approximately a rise of 1 degree
Содержание LABSCI 11L CPVG
Страница 18: ...Page 17 ON OFF On Off LAB048 0018 Near the Power switch...
Страница 85: ...Page 84 1 2...