Page 85
12.6
Cleaning the water outlet strainer
Caution!
Before proceeding, Make sure that the electric cord is
disconnected and there is no pressure or water in the chamber.
Warnings!
1.
The strainer’s cover is HOT
Do not touch the strainer’s cap, mounted on the exhaust line,
during and shortly after operation. Touching the hot strainer’s
cap may cause severe injuries.
2.
If maintenance operation is performed while strainer cap is hot,
use heat resistant gloves to avoid injuries.
1.
Open the strainer cap.
2.
Remove the strainer element.
3.
Rinse the strainer with water, using a brush if necessary.
4.
Reinstall the strainer element.
5.
Close the strainer cap
Cap
Gasket
Strainer
element
Strainer
Housing
Water Strainer