Page 83
Pull off the gasket from the door groove and install the new gasket
referring to the drawings as above points 1, 2 and 3.
Caution!
See drawing below for the right direction of the gasket.
12.5
Checking the Safety Valve
Both safety valves, the chamber safety valve (2) and the generator
safety valve (3), are located on the top rear of the autoclave (see Rear
View).
In order to prevent the safety valve from becoming blocked, it is
necessary to allow the steam pressure to escape through it (every
month).
To avoid injuries begin this check while the autoclave is cold.
1.
For the autoclave safety valve, operate the sterilization cycle; for the
generator safety valve, operate the generator.
2.
Allow a pressure of approximately 200 kPa (29-psi) to build up in the
chamber.
3.
Remove water reservoir cover (1).
4.
Pull the ring of the safety valve using a tool, i.e. screwdriver, hook etc
and lift the safety valve ring for 2 seconds. Be careful not to burn your
hands.
5.
Press the STOP key to abort operation, and allow the steam to exhaust
from chamber.
6.
Wait until pressure goes down to zero, only then can the door be
opened.