Page 84
10.9
Start Cycle by Clock mode
•
This mode enables the operator to define the time of the beginning of
the cycle. The maximum possible delay is 24 hours. For further
information see sec. 6.3.2.
“Start Cycle by Clock”.
10.10
Moving the Autoclave
1.
Disconnect the power supply cord.
2.
Disconnect the water and drain hoses.
3.
Disconnect the compressed air hoses (if applicable).
4.
Drain the water from the chamber.
To avoid injuries, Moving the autoclave should be done by using a
forklift.
Caution:
Before moving the autoclave, verify that the electrical, air and
water connections have been disconnected, and there is no
pressure in the chamber.
Do not drop this device!
10.11
Loading and Unloading the Device
10.11.1
Safety
Protective equipment and clothes and other safety instructions should
be implemented in accordance with local and national regulations
and/or rules!
For proper sterilization - Do not overload the chamber. Only
autoclavable products shall be used; please refer to the manufacturer
instructions for sterilization of unknown materials or instruments.
10.11.2
Loading
Correct loading of the autoclave is essential to successful sterilizing for
several reasons. Efficient air removal from the chamber and the load
will permit effective steam penetration and saturation and allow proper
drainage of condensate. Additionally, correct loading will prevent
damage to packs and their contents and maximize efficient use of the
sterilizer.
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...