22
6.14
Replacing the thermostat B10
Caution:
Before starting, be sure that the electric cord is disconnected from the
power source and that there is no pressure in the autoclave chamber.
1. Remove the autoclave cover (see para. 6.2
“
Dismantling the Outer
Covers of the Autoclave
”
).
2. Unscrew the nut (1) connecting the pressure pipe (the pipe
connecting the thermostat to the chamber).
3. Remove the isolating cover.
4. Remove the thermostat knob.
5. Unscrew the 2 screws connecting the thermostat to the panel
(located under the thermostat knob).
6. Unscrew the nuts connecting the wires to the thermostat (2).
7. Remove the thermostat and replace it with a new one.
8. Reconnect the electrical wires.
9. Reassemble the thermostat to the panel.
10. Reassemble the knob and the pressure pipe.
11. Re-assemble the isolating and the autoclave cover.
12. Test and calibrate the pressure switch.
13. Test all the autoclave cycles and verify it operates correctly.
1
2
Содержание 1730 M
Страница 2: ......
Страница 11: ...9 REAR VIEW...
Страница 34: ...32 GENERAL VIEW OF VESSEL DOOR AND ACCESSORIES Door Closing device Chamber...
Страница 35: ...33 AUTOCLAVE COVER...
Страница 37: ...35 MULTI PURPOSE VALVE ASSEMBLY Model 1730 Models 2340 2540 3140 3850 3870...
Страница 46: ...44 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 1730M MK...
Страница 47: ...45 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 2340 2540 M MK...
Страница 48: ...46 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODEL 3140 M...
Страница 49: ...47 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODELS 3850 M...
Страница 50: ...48 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 3870 M...
Страница 51: ...49 PIPING DIAGRAM TABLE TOP AUTOCLAVE MODELS M AND MK...