Turtle Tough TT-CL2 Скачать руководство пользователя страница 5

TT-CL

2

  

Free Chlorine Sensor 

2.0

 INSTALLATION

As with all instrumentation the installation and commissioning of this instrument is crucial to its safe and accurate 
function. This instrument must be used only for its purpose as outlined in this manual and must be installed and 
commissioned in accordance with this manual and by trained and qualified personnel.

2.1

 Site Selection

Please choose a suitable location for the installation of the probe and the electronics. The choice of installation 
point on any site is a compromise and is best undertaken by experienced installation personnel. The following is 
a list of the factors that need to be taken into consideration. This list is not intended as a checklist neither is it 
implied that the list is complete.

•  Ensure that the mounting allows access to all serviceable parts.

•  Try to mount the electronics in a position where they are not habitually hosed down in a cleaning process.

•  Consider the length of wiring runs when mounting the instrument.

•  Try to keep the electronics away from substations or other large EMI emitters.

•  Consider whether the sample will be representative and well mixed.

•  Consider sample return points.

•  In a plastic run, with a low conductivity sample consider earthing the sample.

2.2 

Unpacking

Please have a copy of your order with you when you unpack your instrument. All orders are checked when they 
leave the factory. Please double check that you have all the parts that were ordered as soon as you open the box. 
This may include (but not necessarily);

 

Probe 

 

Probe mounting overflow cell. 

 

Electrolyte 

 

Membranes 

 

Abrasive paper (blue) 

 

Manual 

 

CO2 Buffer kit  - option 

 

Acid buffer unit - option

If anything is missing, or damaged, please contact your sales outlet immediately. If the instrument needs to be 
returned for any reason please follow the instructions given in Appendix C of this manual.

TURTLE TOUGH

5

Содержание TT-CL2

Страница 1: ...TURTLE TOUGH OPERATION MANUAL TT CL2 FREE CHLORINE SENSOR Ion Specific Chlorine Analyser turtletoughsensors com...

Страница 2: ...ce 11 4 0 3 1 Routine maintenance 11 4 0 3 2 Diagnostic maintenance 11 Storage 11 Spares 11 Single flowcell 11 4 0 6 1 Insertion 12 Double flowcell 13 COMMISSIONING 13 TROUBLESHOOTING GUIDE 13 Further...

Страница 3: ...ill only be necessary to reduce the pH to below 6 5 This is best done using the CO2 Buffer accessory The probes are available for a variety of ranges which are specified at the time of purchase The se...

Страница 4: ...l using DPD or Automatic Autocal Housing material PVC silicone polycarbonate stainless steel perspex Dimensions Diameter approx 25mm 1 length 175mm 7 Maintenance intervals Calibration On control appli...

Страница 5: ...mount the electronics in a position where they are not habitually hosed down in a cleaning process Consider the length of wiring runs when mounting the instrument Try to keep the electronics away from...

Страница 6: ...flow cell single or double with its backplate 9mm thick please refer to the drawings below 2 4 Water connections The incoming water sample connection is to the flowcell lower barb or push fitting In...

Страница 7: ...ole is blocked off when the membrane cap is removed then the membrane may rupture i e develop a tiny tear and the probe may not produce accurate readings When the membrane cap is filled with electroly...

Страница 8: ...electronics in the sensor The measuring signal is independent from the temperature of the measuring water due to an integrated temperature compensator 4 0 1 Installation Please refer to Site Selectio...

Страница 9: ...lectrolyte Be careful so that there are hardly any bubbles Then replace it onto the bench NOTE Read the electrolyte MSDS sheet Fill up the G holder with electrolyte Be careful that there are no bubble...

Страница 10: ...rolyte or electrolyte on your skin or in your eye should be washed off immediately with water The electrolyte contains potassium halide Check that the membrane cap is completely screwed in up to the s...

Страница 11: ...mbrane cap When putting the probe back into use after storage the electrode tip must be cleaned with the special abrasive paper supplied see 4 0 3 2 and a new membrane cap and G holder must be used fo...

Страница 12: ...TT CL2 Free Chlorine Sensor 12 TURTLE TOUGH 4 0 6 1 Insertion There should be 25mm 1 of clear space between the tip of the sensor and the conical base of the flowcell tube...

Страница 13: ...2 Check the sensor for damage 3 Make up the sensor 4 Ensure that water flows through the flowcell and over each sensor s measuring end 5 Calibration After the Cl2 sensor has completely polarized from...

Страница 14: ...TT CL2 Free Chlorine Sensor 14 TURTLE TOUGH Appendix A General Layout Drawing Flowcell...

Страница 15: ...of correctly you will help prevent potential negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of this product The recycling of...

Страница 16: ...Serial number Medium concentration Temperature Pressure Conductivity Viscosity Cleaned with Please mark the appropriate warnings Radioactive Explosive Caustic Poisonous Harmful to health Biologically...

Страница 17: ...age or failure caused by A Improper installation operation or maintenance of the equipment B Abnormal wear and tear on moving parts caused by some processes C Acts of nature i e lightning flooding etc...

Страница 18: ...TT CL2 Free Chlorine Sensor 18 Appendix E Safety Data Sheet Electrolyte ECS2 1 GEL Electrolyte FC1 for Membrane Free Cl2 Sensor Manual date 12 1 2011 Page 25 of 29...

Страница 19: ...TT CL2 Free Chlorine Sensor 19 Appendix E IPM19 HaloSense Free Chlorine Probe...

Страница 20: ...TT CL2 Free Chlorine Sensor IPM19 HaloSense Free Chlorine Probe 20...

Страница 21: ...value measured perform the probe maintenance procedure Calibration Failed Please wait until process is stable for calibration would be displayed if calibration fails Finish to save and exit to the men...

Отзывы: