Turtle Tough TT-CL2 Скачать руководство пользователя страница 11

TT-CL

2

  

Free Chlorine Sensor 

11

TURTLE TOUGH

4.0.3

 Maintenance

Caution: The coating of the electrode finger must not be cleaned with emery cloth or other polishing agents! 

Do not unscrew the metal end ring of the membrane cap, as this will destroy the membrane.

4.0.3.1

 Routine maintenance

The electrolyte should be changed every three months, although it may be possible to extend the period if low 
levels of chlorine are measured. It may be necessary to use a new membrane cap when changing the electrolyte. 
In any event, the membrane cap should be changed every 12 months. If your unit has been ordered with the auto 
calibrator then the calibration reagent should be changed every six months.

4.0.3.2

 Diagnostic maintenance

4.0.4  

Storage

To  store  the  probe  the  membrane  cap  is  unscrewed  (note  previous  warning). The  membrane  G-holder  must 
be removed and discarded. The membrane cap and electrode finger should be rinsed in clean water and dried 
in a place free of dust with a soft paper towel (do not damage the electrode areas). The dry membrane cap is 
then loosely screwed onto the electrode shaft. The membrane must not rest against the measuring electrode.  
Do not fully tighten the membrane cap.

When putting the probe back into use after storage, the electrode tip must be cleaned with the special abrasive 
paper supplied (see 4.0.3.2) and a new membrane cap and G-holder must be used (follow 4.0.2).

4.0.5

  Spares 

Free Chlorine Membrane cap  

M48 G 

Free Chlorine Electrolyte  

ECS2.1/GEL

Please contact us regarding any other spares.

4.0.6

  Single flowcell

The o-ring is first inserted in the 1” opening of the adaptor followed by the PVC shim. Then the 1” PVC threaded 
end cap is screwed in loosely. The probe is inserted into this prepared probe housing. Then the probe is fixed 
in place by tightening (hand tight only) the threaded end cap, so the probe cannot move away from its position.

 

If  calibration  is  impossible  due  to  unstable  or  low  output,  
remove the sensor, displace the vent seal so that the vent hole 
opening is free. The membrane cap can then be unscrewed. The  
electrode finger is cleaned with a clean, dry, soft paper towel. 
With the special abrasive paper supplied (blue) just the tip of 
the dry electrode (the working electrode) is cleaned. Place the  
special abrasive paper on a paper towel (on a flat, dry surface), hold 
it at the corners and rub the electrode tip of the perpendicularly 
held probe two or three times across the abrasive paper. Then 
refill with electrolyte (see section 4.0.2). If necessary, use a new  
membrane cap. 

Recommendation:

 Change the electrolyte every 3 months.

Содержание TT-CL2

Страница 1: ...TURTLE TOUGH OPERATION MANUAL TT CL2 FREE CHLORINE SENSOR Ion Specific Chlorine Analyser turtletoughsensors com...

Страница 2: ...ce 11 4 0 3 1 Routine maintenance 11 4 0 3 2 Diagnostic maintenance 11 Storage 11 Spares 11 Single flowcell 11 4 0 6 1 Insertion 12 Double flowcell 13 COMMISSIONING 13 TROUBLESHOOTING GUIDE 13 Further...

Страница 3: ...ill only be necessary to reduce the pH to below 6 5 This is best done using the CO2 Buffer accessory The probes are available for a variety of ranges which are specified at the time of purchase The se...

Страница 4: ...l using DPD or Automatic Autocal Housing material PVC silicone polycarbonate stainless steel perspex Dimensions Diameter approx 25mm 1 length 175mm 7 Maintenance intervals Calibration On control appli...

Страница 5: ...mount the electronics in a position where they are not habitually hosed down in a cleaning process Consider the length of wiring runs when mounting the instrument Try to keep the electronics away from...

Страница 6: ...flow cell single or double with its backplate 9mm thick please refer to the drawings below 2 4 Water connections The incoming water sample connection is to the flowcell lower barb or push fitting In...

Страница 7: ...ole is blocked off when the membrane cap is removed then the membrane may rupture i e develop a tiny tear and the probe may not produce accurate readings When the membrane cap is filled with electroly...

Страница 8: ...electronics in the sensor The measuring signal is independent from the temperature of the measuring water due to an integrated temperature compensator 4 0 1 Installation Please refer to Site Selectio...

Страница 9: ...lectrolyte Be careful so that there are hardly any bubbles Then replace it onto the bench NOTE Read the electrolyte MSDS sheet Fill up the G holder with electrolyte Be careful that there are no bubble...

Страница 10: ...rolyte or electrolyte on your skin or in your eye should be washed off immediately with water The electrolyte contains potassium halide Check that the membrane cap is completely screwed in up to the s...

Страница 11: ...mbrane cap When putting the probe back into use after storage the electrode tip must be cleaned with the special abrasive paper supplied see 4 0 3 2 and a new membrane cap and G holder must be used fo...

Страница 12: ...TT CL2 Free Chlorine Sensor 12 TURTLE TOUGH 4 0 6 1 Insertion There should be 25mm 1 of clear space between the tip of the sensor and the conical base of the flowcell tube...

Страница 13: ...2 Check the sensor for damage 3 Make up the sensor 4 Ensure that water flows through the flowcell and over each sensor s measuring end 5 Calibration After the Cl2 sensor has completely polarized from...

Страница 14: ...TT CL2 Free Chlorine Sensor 14 TURTLE TOUGH Appendix A General Layout Drawing Flowcell...

Страница 15: ...of correctly you will help prevent potential negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of this product The recycling of...

Страница 16: ...Serial number Medium concentration Temperature Pressure Conductivity Viscosity Cleaned with Please mark the appropriate warnings Radioactive Explosive Caustic Poisonous Harmful to health Biologically...

Страница 17: ...age or failure caused by A Improper installation operation or maintenance of the equipment B Abnormal wear and tear on moving parts caused by some processes C Acts of nature i e lightning flooding etc...

Страница 18: ...TT CL2 Free Chlorine Sensor 18 Appendix E Safety Data Sheet Electrolyte ECS2 1 GEL Electrolyte FC1 for Membrane Free Cl2 Sensor Manual date 12 1 2011 Page 25 of 29...

Страница 19: ...TT CL2 Free Chlorine Sensor 19 Appendix E IPM19 HaloSense Free Chlorine Probe...

Страница 20: ...TT CL2 Free Chlorine Sensor IPM19 HaloSense Free Chlorine Probe 20...

Страница 21: ...value measured perform the probe maintenance procedure Calibration Failed Please wait until process is stable for calibration would be displayed if calibration fails Finish to save and exit to the men...

Отзывы: