Turtle Tough TT-CL2 Скачать руководство пользователя страница 3

TT-CL

2

  

Free Chlorine Sensor 

SAFETY PRECAUTIONS                                                   

•  Before attempting to unpack, set up, or operate this instrument, please read this entire manual.
•  Make certain the unit is disconnected from the power source before attempting to  service or remove any 

component.

Make certain the unit is disconnected from other sources of force or pressure (for example, pneumatic or hydrau-
lic), before attempting to service or remove any component.

•  Failure to follow these precautions could result in personal injury and damage to the equipment.

 

1.

 THEORY OF OPERATION

The free chlorine sensor has been designed to measure the concentration of dissolved free residual chlorine. 
It  does  this  using  a  Potentiastatic  3-electrode  chronoamperometric  sensor  insulated  from  the  water  supply 
by a hydrophilic membrane. The free chlorine diffuses through the membrane and is reduced at the cathode  
generating  a  small  current  that  is  proportional  to  the  concentration  of  free  chlorine.  In  potable,  process,  or  
swimming pool water this means HOCl (hypochlorous acid) and OCl- (hypochlorite ion). The relative amount of 
these two species is dependent on the pH of the solution. At low pH’s the predominant species will be HOCl. 
At higher pH's the predominant species will be OCl-. Traditional amperometric measurement systems need to 
be buffered to an exact pH in order that changes in the pH of the sample water do not affect the Free Chlorine 
measurement. The probe supplied for Free Chlorine measurement with this instrument is not affected by changes 
in pH to the same degree. This means that on many plants it will not be necessary to buffer the sample water at 
all. If it is necessary to buffer the sample water, because the pH is both high and variable it will only be necessary 
to reduce the pH to below 6.5. This is best done using the CO2 Buffer accessory. The probes are available for a 
variety of ranges which are specified at the time of purchase.

The sensor incorporates automatic temperature compensation (ATC).

3

TURTLE TOUGH

IPM19  HaloSense Free Chlorine Probe

Manual date 12/1/2011

Page 4 of 29

1.0

Theory of Operation

The free chlorine sensor has been designed to measure the concentration of dissolved free residual 
chlorine. It does this using a Potentiastatic 3-electrode chronoamperometric sensor insulated from the water
supply by a hydrophilic membrane. The free chlorine diffuses through the membrane and is reduced at the 
cathode generating a small current that is proportional to the concentration of free chlorine. In potable, 
process, or swimming pool water this means HOCl (hypochlorous acid) and OCl

-

(hypochlorite ion). The 

relative amount of these two species is dependent on the pH of the solution. At low pH’s the predominant 
species will be HOCl. At higher pH's the predominant species will be OCl

-

. Traditional amperometric 

measurement systems need to be buffered to an exact pH in order that changes in the pH of the sample 
water do not affect the Free Chlorine measurement. The probe supplied for Free Chlorine measurement 
with this instrument is not affected by changes in pH to the same degree. This means that on many plants it 
will not be necessary to buffer the sample water at all. If it is necessary to buffer the sample water, because 
the pH is both high and variable it will only be necessary to reduce the pH to below 6.5. This is best done 
using the 

CO

2

Buffer

accessory. The probes are available for a variety of ranges which are specified at the 

time of purchase.
The sensor incorporates automatic temperature compensation (ATC).

Typical Probe Response to pH 

(Unbuffered)

0

20

40

60

80

100

120

6

6.5

7

7.5

8

8.5

9

9.5

10

Chlorine Probe Response

Typical Amperometric Probe Response

Содержание TT-CL2

Страница 1: ...TURTLE TOUGH OPERATION MANUAL TT CL2 FREE CHLORINE SENSOR Ion Specific Chlorine Analyser turtletoughsensors com...

Страница 2: ...ce 11 4 0 3 1 Routine maintenance 11 4 0 3 2 Diagnostic maintenance 11 Storage 11 Spares 11 Single flowcell 11 4 0 6 1 Insertion 12 Double flowcell 13 COMMISSIONING 13 TROUBLESHOOTING GUIDE 13 Further...

Страница 3: ...ill only be necessary to reduce the pH to below 6 5 This is best done using the CO2 Buffer accessory The probes are available for a variety of ranges which are specified at the time of purchase The se...

Страница 4: ...l using DPD or Automatic Autocal Housing material PVC silicone polycarbonate stainless steel perspex Dimensions Diameter approx 25mm 1 length 175mm 7 Maintenance intervals Calibration On control appli...

Страница 5: ...mount the electronics in a position where they are not habitually hosed down in a cleaning process Consider the length of wiring runs when mounting the instrument Try to keep the electronics away from...

Страница 6: ...flow cell single or double with its backplate 9mm thick please refer to the drawings below 2 4 Water connections The incoming water sample connection is to the flowcell lower barb or push fitting In...

Страница 7: ...ole is blocked off when the membrane cap is removed then the membrane may rupture i e develop a tiny tear and the probe may not produce accurate readings When the membrane cap is filled with electroly...

Страница 8: ...electronics in the sensor The measuring signal is independent from the temperature of the measuring water due to an integrated temperature compensator 4 0 1 Installation Please refer to Site Selectio...

Страница 9: ...lectrolyte Be careful so that there are hardly any bubbles Then replace it onto the bench NOTE Read the electrolyte MSDS sheet Fill up the G holder with electrolyte Be careful that there are no bubble...

Страница 10: ...rolyte or electrolyte on your skin or in your eye should be washed off immediately with water The electrolyte contains potassium halide Check that the membrane cap is completely screwed in up to the s...

Страница 11: ...mbrane cap When putting the probe back into use after storage the electrode tip must be cleaned with the special abrasive paper supplied see 4 0 3 2 and a new membrane cap and G holder must be used fo...

Страница 12: ...TT CL2 Free Chlorine Sensor 12 TURTLE TOUGH 4 0 6 1 Insertion There should be 25mm 1 of clear space between the tip of the sensor and the conical base of the flowcell tube...

Страница 13: ...2 Check the sensor for damage 3 Make up the sensor 4 Ensure that water flows through the flowcell and over each sensor s measuring end 5 Calibration After the Cl2 sensor has completely polarized from...

Страница 14: ...TT CL2 Free Chlorine Sensor 14 TURTLE TOUGH Appendix A General Layout Drawing Flowcell...

Страница 15: ...of correctly you will help prevent potential negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of this product The recycling of...

Страница 16: ...Serial number Medium concentration Temperature Pressure Conductivity Viscosity Cleaned with Please mark the appropriate warnings Radioactive Explosive Caustic Poisonous Harmful to health Biologically...

Страница 17: ...age or failure caused by A Improper installation operation or maintenance of the equipment B Abnormal wear and tear on moving parts caused by some processes C Acts of nature i e lightning flooding etc...

Страница 18: ...TT CL2 Free Chlorine Sensor 18 Appendix E Safety Data Sheet Electrolyte ECS2 1 GEL Electrolyte FC1 for Membrane Free Cl2 Sensor Manual date 12 1 2011 Page 25 of 29...

Страница 19: ...TT CL2 Free Chlorine Sensor 19 Appendix E IPM19 HaloSense Free Chlorine Probe...

Страница 20: ...TT CL2 Free Chlorine Sensor IPM19 HaloSense Free Chlorine Probe 20...

Страница 21: ...value measured perform the probe maintenance procedure Calibration Failed Please wait until process is stable for calibration would be displayed if calibration fails Finish to save and exit to the men...

Отзывы: