Turtle Tough CPVC Скачать руководство пользователя страница 12

TURTLE

TOUGH

®

 

12

on either the process side of the power/electrical side and they can even occur through the atmosphere/air. 

 

As such the pH/ORP sensor should be installed as far away as possible from high current carrying equipment such 
as pumps, motors, variable speed drives etc.

Position-Sensitive Sensors

It is recommended that pH Sensors be optimally installed at a 45° angle as this avoids air bubbles being trapped 
at  the  sensor/process  interface  and  ensures  optimal  glass/silver-chloride  interaction.  It  is  also  acceptable  to 

 

install vertically down. pH sensors should never be installed inverted (ie upside down) or below the horizontal 
axis. They must be at least 15° above the horizontal axis.

Position Insensitive Sensors

For ORP and Conductivity, these are both positions insensitive, meaning the sensor can be installed at any angle. 
pH and Dissolved Oxygen on the other hand are not, and must always be in the vertically down position. It is still 
recommended to install these sensors vertically down (tip down) for several good reasons:

• 

Upside down is messy and can result in product spillage and overflow.

• 

It can be unsafe if the process is not isolated correctly.

• 

Product with any solids… solids can settle out and get trapped in the sensor tip and cause excess fouling.

• 

People see an ORP or Conductivity sensor upside…. They then install a pH or DO sensor upside down 
(which is not good).

Should you require any more assistance regarding the installation of our sensors, please contact Turtle Tough.

Содержание CPVC

Страница 1: ...1 MANUAL pH ORP SENSOR Installation Calibration Cleaning Care Maintenance TURTLE TOUGH turtletoughsensors com...

Страница 2: ...es 18 Common offset type issues 18 pH buffer accuracy and stability issues 18 pH buffer during calibration 19 Temperature effects on pH reading 19 Sensor Care and Maintenance 20 Storage 20 Maintenance...

Страница 3: ...eading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety of...

Страница 4: ...Shock Connect only to a suitable isolated hard wired electrical outlet It is recommended that the outlet is protected by an RCD Residual Current Detector or is in any event in compliance with all loca...

Страница 5: ...ling them CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective c...

Страница 6: ...ity while resisting aggressive gas intrusion is what makes our sensors outlast virtually all others in any process conditions Your Turtle Tough sensor is built with our unique and proprietary glass fo...

Страница 7: ...able in 4 main body plastics which include CPVC Chlorinated Polyvinyl Chloride RYTON Poly Phenylene Sulfone RADEL Poly Phenyl Sulfone PEEK Poly Ether Ether Ketone Each body plastic serves a function b...

Страница 8: ...0 2 00 25 40 1 00 38 10 1 50 88 90 3 50 Knurled Section Max 28 4 1 119 Diameter 3 4 NPT Pipe Thread 3 4 NPT Pipe Thread 19 05 75 Measurements U Un ni it ts s m mm m i in nc ch he es s Rev Date of issu...

Страница 9: ...1 1 NPT Pipe Threads 118 62 4 67 Measurements U Un ni it ts s m mm m i in nc ch he es s Rev Date of issue Sheet Code 1 1 04 11 2022 Created by Patrick Voorhoeve Website turtletoughsensors com Product...

Страница 10: ...your Turtle Tough representative immediately If the instrument needs to be returned for any reason please follow the return instructions given in this manual Please dispose of the packing in an envir...

Страница 11: ...ollowing Where possible avoid using front end threaded connections When threading the sensor in out of the process continuously ie each time you clean and calibrate the cable is constantly exposed to...

Страница 12: ...ey must be at least 15 above the horizontal axis Position Insensitive Sensors For ORP and Conductivity these are both positions insensitive meaning the sensor can be installed at any angle pH and Diss...

Страница 13: ...of the sen sor The most common approach so to use sufficient TEFLON tape and an FNPT coupling to create this back sensor seal and then either a rigid or flexible conduit back to the transmitter where...

Страница 14: ...perly running the cable in conduit back to the transmitter The use of Standard or Extreme Water proofing options is VERY STRONGLY recommended although not strictly required for submersible installatio...

Страница 15: ...e Waterproofing TURTLE TOUGH SCALE SHEET SIZE FSCM NO DWG NO REV REVISIONS ZONE REV DESCRIPTION DATE APPROVED A B C D A B C D 1 2 3 4 1 2 3 4 DRAWN BY DATE PRC 8 5 99 BLACK BAIRDLTD 1 OF 1 1 ASSEMBLY...

Страница 16: ...rovide optimal life when connected to a TT MA Online Analyser The connection to the TT pHP is via BNC and banana plug connectors Wiring of non preamp preamp and external preamp sensors to the TT MA ar...

Страница 17: ...then in a pH buffer above pH7 in Slope mode You must exit the Slope mode by using the Mode key after completing the acidic slope below pH7 calibration before entering the alkaline slope above pH7 cali...

Страница 18: ...er all pH buffer calibrations have been completed Ph Buffer Accuracy and Stability Issues It is important to consider that some pH buffers degrade in quality much more quickly than others when left op...

Страница 19: ...account for the change in the pH buffer values due to temperature and so this correction must be done manually Temperature Effects on pH Reading More generally note that temperature compensation for...

Страница 20: ...y need to store your pH or ORP sensor if you don t need to use it right away or you are using it intermittently Please follow the storage steps below to ensure your sensor s longevity STEP 1 Always st...

Страница 21: ...tool IMPORTANT Great care should be taken not to scratch the pH glass or orp element during cleaning of the reference junction Common approved cleaning solutions include 5 15 Hydrochloric Acid For Alk...

Страница 22: ...sor While this question is virtually impossible to answer a Turtle Tough sensor will have up to 10 times the stability of mainstream sensors Most of our customers can extend calibration intervals by 2...

Страница 23: ...ide if the reference junction has been depleted of the ions in the solid state conductive polymer typical for clean water applications Condition the sensor in saturated potassium chloride and or pH 4...

Страница 24: ...irreparably damage the sensor 3 Once most of the fouling is removed pour some of the HCI solution into the beaker and soak the tip of the sensor in the HCI for 5 to 10 minutes Then rinse the tip thor...

Страница 25: ...mount plugs that directly accept the male sensor plug There will be times when you want to extend cable runs or integrate sensors directly with your hardware Turtle Tough make a variety of hardware t...

Страница 26: ...y issues cannot be supported if you are using your cables and you should always have at least one genuine Turtle Tough interface on hand for troubleshooting or testing Please contact Turtle Tough for...

Страница 27: ...m L a Rod top section b Rod bottom section Comes with Tank holder assembly Female fittings Mounting plate Submersible Immersion B Rod bottom Junction Box not included C Tank holder assembly Sensor Ins...

Страница 28: ...Please note that our quick connect plugs my never be cut or removed from the sensor This will immediately void warranty Only use Turtle Tough factory interface leads or panel mount connectors to inter...

Страница 29: ...e faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide based on simila...

Страница 30: ...and plugs must NOT be cut removed or modified in any way Sensors contain sensitive internal electronics and our cables and connectors are designed to protect the integrity of these components Any mod...

Страница 31: ...aterial plus the name and phone number of a contact person that can be reached for further information Clean Statement returned goods that have been in contact with process fluids must be decontaminat...

Страница 32: ...support is limited to the information provided in our field implementation notes and guidelines The product is fully supported for installation requirements field implementation and programming proto...

Отзывы: