2. Instrument Operation and Calibration
2.1 Button Identification and Function
CAL
– Is used to calibrate the instrument. When selected it will prompt the user to insert
blank. After blank is read, it will prompt user to insert CAL solution. After CAL solution
is read, the screen will display “Calibration Done”.
NOTE: SEE APPENDIX C BEFORE CALIBRATING THE AquaFlash.
CAL VAL
– Is used to assign a value to the calibration solution used when calibrating
the
Aqua
Flash. When selected it will prompt the user to set a calibration value from 1 to
50. After the value is set, the user will press enter to save the setting.
READ
– Begins the sample analysis procedure detailed in section 2.4.
ON/OFF
– Turns the instrument ON or OFF.
ENTER
–Allows users to proceed to the next screen/menu.
DATA
– When the DATA button is pressed users can choose between either
sending/clearing data or setting date/time. Use the
keys to choose between the two
options; a marker will indicate the selection made.
If SEND/CLEAR DATA is selected, press ENTER and a submenu will display. Users
can choose, using
keys, to either send data to a connected computer, see section
2.6 or clear logged data; a marker will indicate the selection made.
o
If
SEND DATA
is selected and ENTER is pressed, the instrument will send all logged
data out as an ASCII file, see section 2.6.
o
If
CLEAR DATA
is selected, the following message will display;
“ENTER TO
CONFIRM CLEAR DATA
”; when ENTER is pressed all logged data will be deleted.