VI. Calibration of the GloRunner
Overview
Calibration of the GloRunner
is possible if you purchased the GloRunner
Standardization Light Plate. The calibration procedure is password protected and
should only be done after contacting Turner BioSystems for the password. This
calibration is a process that normalizes the detectors and is not usually required.
If it is determined that detector normalization is required, choose “Utilities” on
the pull down menu. Next select “Calibrate Using Light Plate”. You will see the
following pop-up menu:
Figure 20
Procedure
1.
Type in the password and click on the “Enter New Password” button. You
will be prompted with the following pop-up button.
Figure 21
2.
Turn on the light plate by briefly pressing on the “Start” button. Insert it
into the sample tray with the proper plate placement. (Proper plate
placement is the alignment notch corner of the light plate to the right rear
of the sample tray. See Figure 22). Click “OK”.
Figure 22