Turf-Tec ECM1-N Скачать руководство пользователя страница 3

 

Page 3 - 

www.turf-tec.com

  

 
See Appendix 1 for a list of preferred EC values for some common turfgrass varieties. 
Typically, younger seedling of turfgrass or newly planted sprigs will require lower EC 
than mature plants. The stainless-steel soil EC probe is designed to be inserted directly 
into soil. The sensing surface is composed of 2 pairs of electrodes on the probe tip.  
Additionally, the probe is narrow in diameter so it can be used effectively on golf greens 
and sports fields. 

  

Because EC readings are affected by moisture content, it is important that soil moisture 
content does not differ significantly between readings. An easy way to achieve this  
condition is by taking measurements approximately 30 to 60 minutes after an irrigation. 
This should ensure the soil moisture level has approximately reached field capacity. The 
probe should be inserted in the root zone. The measurement region is at the tip of the 
probe. For turf, the root zone is approximately 2” - 4”. For vegetables and small plants, 
this is about 8” - 12”. Wait until the meter reading stabilizes before withdrawing the 
probe. Taking several measurements will allow a representative average to be computed. 
 

Important:  Do not touch the sensor tip with your fingers.  The oils on the skin will 
affect the probe’s measurement accuracy. 
 

Meter Settings 

Accessing the Setup Screen 

 

Power on the tester. Press the MENU/

▼ button. The meter will briefly display the 

Key Info screen before transitioning to the Setup screen. Follow the instructions for 
the specific parameter that is to be modified. Pressing the Cal/Esc button will return 
you to the measurement screen.  

 

Measurement Parameter 

 

1. Press MENU/

▼ to scroll to Measurement option. 2. Press HOLD. The display 

shows Cond, TDS, and Salinity. 

 

3. Use the MENU/

▼ button to toggle between Cond, TDS, and Salinity. Press HOLD 

to select desired option. 

 

4. The display shows the selected parameter with a checkmark and returns to the 
Setup screen.  

 

TDS Factor 

 

1. Press MENU/

▼ to scroll to Settings option. 

 

2. Press HOLD. The display shows TDS factor and Backlight. 

 

3. Use the MENU/

▼ button to scroll to the TDS Factor option 

 

4.Press HOLD. Current adjustment factor will be displayed. 

 

5.Adjust the setting with the MENU/

▼ button. Note that the setting can only be 

adjusted down. The value will flip to 1.00 after you scroll past 0.40. 

 

6. Press HOLD to confirm the TDS factor setting. The display shows the TDS factor 
with a check mark then returns to the Setup screen.  

 
 
 
 
 

Содержание ECM1-N

Страница 1: ...paration You can test different levels in the soil profile with the eight inch long probe You can also use the probe to fine tune your fertility program Measurement can be made directly in the soil wi...

Страница 2: ...Use this portable EC meter and probe to measure salinity directly in the soil without tedious soil sampling and preparation Turfgrass managers can monitor for high salt levels on golf course greens an...

Страница 3: ...s will allow a representative average to be computed Important Do not touch the sensor tip with your fingers The oils on the skin will affect the probe s measurement accuracy Meter Settings Accessing...

Страница 4: ...e upper display shows the current measured temperature reading based on the factory default calibration 5 Dip the tester into a solution of known temperature and allow time for the built in temperatur...

Страница 5: ...rior to measurement Prior to calibration the tip of the probe should be cleaned with alcohol The tester will retain one calibration value in each mode conductivity TDS salinity when the instrument is...

Страница 6: ...ments at different soil depths you can determine where the salts are concentrated in the soil profile Soil moisture content will significantly influence the measured EC value To ensure accurate measur...

Страница 7: ...if needed The sample should have the consistency of a paste with slightly more water than if the media was in a pot and fully irrigated Press On Off button to power up the meter Insert the probe tip...

Страница 8: ...ter of the meter opening Be sure that the small O ring sits at the base of the large plastic assembly 6 While holding the assembly in place hand tighten the nut to the strain relief 7 Connect and tigh...

Страница 9: ...resolution 0 1 C 0 1 F Temperature accuracy From 0 to 50 C 0 5 C 0 9 F 1 LSD from 50 to 60 C 1 0 C 1 8 F 1 LSD General Display Graphics Dot matrix 80 x 100 pixel Auto off 8 5 minutes from last key pre...

Страница 10: ...0 3 1 6 1 3 4 3 2 3 7 0 0 4 1 9 1 4 4 6 2 4 7 3 0 5 2 2 1 5 4 9 2 5 7 6 0 6 2 4 1 6 5 1 2 6 7 8 0 7 2 7 1 7 5 4 2 7 8 1 0 8 3 0 1 8 5 7 2 8 8 4 0 9 3 2 1 9 5 9 2 9 8 6 1 0 3 5 2 0 6 2 3 0 8 9 Table 1...

Страница 11: ...transportation prepaid to Turf Tec International after contacting Turf Tec International and receiving a return authorization number Please include a dated proof of purchase with your tool ANY IMPLIE...

Отзывы: