Turbines Incorporated 

Cryogenic Flow Delivery System 

CDS1000 Manual (Rev E)

 

 

 

CDS1000 Page 

13

 

Specifications 

Hardware: 

 

Amplifier Input: 

 

Frequency range: 0-2500 Hz 

 

Sensitivity: 

 

15mv RMS @ 10 Hz (min) 

 

100mv RMS @ 500 Hz (min) 

 

400mv RMS @ 2.5 KHz (min) 

 

Impedance: Approx. 10k ohms 

 

Noise immunity: High frequency roll off, FCO @ 2.5 KHz 

 

Pickup coil failure detection: 500µA excitation to read coil resistance. 3250 ohms full scale. 

 

Sensor Inputs: 

 

Temperature probe: 

 

1000 ohm platinum 2 wire RTD probe; .00385

/

 ºC European din curve 

 

8mA constant current sensor excitation, 200 ohms full scale 

 

Ratio metric and chopper stabilized to avoid drift 

 

Power: Continuous monitoring of main power, backup battery, & sensors. 

 

Outputs: 

 

Pump Control: (1) Form C 2A relay 

 

Alarm: (1) Form C 2A relay 

 

Communications: RS232 & Bluetooth 

 

Indicators: Front panel LED’s Power (blue), Signal (white), Alarm (red) 

 

Keypad: 16 button keypad, 4 softkey (A, B, C, D), ON/OFF button. Oil resistant, sealed 

 

Display:  

 

240x128 full graphic LCD with back light 

 

Auto LCD bias control, self-regulating non-sweating temperature activated heater 

 

Power: 

 

External: 9-26 Vdc, self-resetting fuse 

 

0.6A with heater OFF 

 

1.6A with heater ON 

 

Transient protection 

 

Internal: Rechargeable NiMH battery pack. Maintains power for up to 10 seconds after power loss. 
Typical service life: up to 5 years 

 

Physical & Environmental: 

 

Operating temperature: -25 to +50ºC 

 

Humidity: 5-95% non-condensing 

 

Sealed, shock vibration tested enclosure 

 

Dimensions: 9.75(L) X 7.50 (H) X 3.75 (W) 

 

Weight: 4 lbs 

 

Miscellaneous: Real time clock with daylight savings time adjustment, dual 8 MHz CPU's, 1M Flash/64K 
RAM, Watchdog timer, 8 channel 12 bit Hi-speed ADC. Buffered serial port, FIFO CPU communication. 

Software: 

 

System Integrity: Built-in diagnostics for all hardware. Power-on self test Constant system monitoring of 
inputs and sensors 

 

Sensor Resolution 

 

RTD - 0.04 ohms approx 

 

Loop- 0.006mA approx. 

 

Math Processing: Internal floating point (FP), 15 digits of precision. Calculations carried out in FP for max 
precision. 

 

Overall Error: 0.001% max (excludes uncertainty of temperature & pressure probe) 

 

Power Management: CPU extinguishes backlight when operating on internal battery. Auto shutdown after 
10 sec power loss 

Содержание CDS1000

Страница 1: ...Box 933 15935 US Hwy 283 N Altus OK 73522 0933 Seneca 864 882 4544 http turbinesincorporated com Manual part number 900484 Revision E Copyright 2013 2016 Turbines Inc All rights reserved Information c...

Страница 2: ...inted in the USA All Rights Reserved SAFETY INSTRUCTIONS The following instructions must be observed Every effort has been made to design and manufacture this instrument to be safe for its intended us...

Страница 3: ...tallation 7 Operation 8 Programming 9 Specifications 13 Hardware 13 Software 13 Troubleshooting 14 Illustrations 15 Wiring 17 Return Policy 18 All Returns 18 Warranty Claim 18 Non Warranty Repair 18 E...

Страница 4: ...ally exclude damage that has resulted from improper installation of improper maintenance of unauthorized change or modification to improper operation of and or misuse or abuse of the subject equipment...

Страница 5: ...DS1000 The displayed total is corrected for temperature by sensing the resistance of the RTD temperature probe Delivery information consisting of 17 selectable parameters is transmitted via Bluetooth...

Страница 6: ...s the normal display for this mode The Detail screen is also accessible in this mode without a password Programming Mode The Programming mode and associated menu s permit the CDS1000 to be configured...

Страница 7: ...b Use at least 4 20 bolts with lock washers and or lock nuts 4 Install cables and connect to the rear panel Do not over tighten Connectors should be hand tight a Route cables and tie wrap excess so th...

Страница 8: ...ss to PRINT only if printer is enabled To View uncompensated total Press to RESET totalizer to ZERO flow must be stopped Press to see unit DETAILS list and perform Battery Test Indicates Pump Cooldown...

Страница 9: ...must be removed 1 Calibration 1 Average K factor 2 Linearizer Table 3 Linearizer Press 0 OFF 1 ON 2 Configuration 1 LOX 2 LIN 3 LAR 4 LCO2 5 LN2O 6 CCO2 7 MAPP 8 LPG Type value press ENT Not normally...

Страница 10: ...Type value press ENT max 9999 Default 1234 for Turbines Bluetooth Printer Next Page Next Page 2 Totalize during alarm 3 Pump cooldown timer 4 Pump cooldown delay 5 Pump ON during alarm 6 Count while...

Страница 11: ...rs 4 Print Example 005 19 Apr 11 14 29 Lin table 004 19 Apr 11 10 51 Factor 158 3444 003 16 Apr 11 15 54 Lin table 002 15 Apr 11 11 54 Factor 149 853 001 12 Mar 11 13 11 Lin table Press any key to con...

Страница 12: ...port on the rear of the unit labeled OUTPUT The print timeout controls when the ticket totals are adjusted for a new ticket When the timer expires the value for start is set to the last value for stop...

Страница 13: ...control self regulating non sweating temperature activated heater Power External 9 26 Vdc self resetting fuse 0 6A with heater OFF 1 6A with heater ON Transient protection Internal Rechargeable NiMH...

Страница 14: ...from turbine meter Reconnect signal cable to coil b With a screw driver rapidly scratch coil below the threads on the bottom surface c Observe CDS1000 counts or signal light slightly illuminates or f...

Страница 15: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 15 Illustrations...

Страница 16: ...Turbines Incorporated Cryogenic Flow Delivery System CDS1000 Manual Rev E CDS1000 Page 16...

Страница 17: ...n Color Function Color 1 Pressure Input 4 20mA Brown Brown 1 Pump NC Brown 1 RTS N C 2 Pressure Input 4 20mA White White 2 Pump C White 2 CTS N C 3 Internal GND Blue Blue 3 Pump NO Blue 3 RX Red 4 Rat...

Страница 18: ...anty Repair If a Product is returned for repair Turbines R D LLC will provide an estimate of the repair charges on a time and materials basis and the time required to make the repair If any repair est...

Страница 19: ...rompted depress Enter and then ESC this places the CDS1000 in the prove mode operating screen The Prove mode allows repeated access to adjust the Kfactor while the cooldown timer flow inhibit of the r...

Страница 20: ...e and remove the PEM With the CDS1000 powered up normally depress Setup and if not already in Prove mode enter your password Select Avg Kfactor by depressing 1 Example A new Kfactor is calculated to b...

Страница 21: ...t page Power Off 11 12 2012 B Delete duplicate programming on Pg11 12 21 2012 C Incorporate Pressure PCO control feature and menu items 06 11 2014 D Added missing menu item 3 2 7 Pump off flow rate er...

Отзывы: