Turbines Incorporated
Cryogenic Flow Delivery System
CDS1000 Manual (Rev E)
CDS1000 Page
11
3) Change system settings (continued)
0) Goto Page 2
Definitions…
Low Power Low input power below 8.75v
Low Batt Backup battery voltage below 8.5v
Coil Short P/U Coil resistance below 250 ohms
Coil Open P/U Coil resistance above 2500 ohms
RTD Short Probe resistance below 100 ohms
RTD Open Probe resistance above 2000 ohms
Tloop Short Temperature loop current above 35mA
Tloop Open Temperature loop current below 4mA
Ploop Short Pressure loop current above 35mA
Ploop Open Pressure loop current below 4mA
(Press any key to continue)
1) View audit trail
7) System alarm log
9) N/A
8) System
Example…
Turbine hrs: 10.9
Pump hrs: 15
System hours: 25.6
Power cycles: 55
Backup Batt sec: 125
Last System PM: 17-Apr-11
Last Turbine PM: 17-Apr-11
Last Pump PM: 17-Apr-11
(Press any key to continue
)
2) Configuration event log
3) Event counters
4) Print
Example…
005 19-Apr-11 14:29 -Lin. table
004 19-Apr-11 10:51 -Factor =158.3444
003 16-Apr-11 15:54 -Lin. table
002 15-Apr-11 11:54 -Factor =149.853
001 12-Mar-11 13:11 -Lin. table
(Press any key to continue)
Example…
004 19-Apr-11 11:00 -DP =0
003 19-Apr-11 10:51 -Product = LOX
-DP =1
002 16-Apr-11 15:54 -Product = LIN
001 15-Apr-11 11:54 -Product = LAR
(Press any key to continue)
Example…
Calibration Counter - 5
Configuration counter - 4
(Press any key to continue)
Must connect a bluetooth or RS-232 printer
1) Calibration event log
Next
Page
Next
Page
* Default Value
5) Maintenance
Enter setting “DDMMYY” press “ENT”
1) System maintenance due
2) System maintenance just performed
3) Turbines hours until maintenance
4) Turbine maintenance just performed
5) Pump hours until maintenance
6) Pump maintenance just performed
Press “0” NO*, “1” YES
Enter setting in hours, press “ENT”
(max 9999)
Press “0” NO*, “1” YES
Enter setting in hours, press “ENT”
(max 9999)
Press “0” NO*, “1” YES
6) Printer Type
Press 1-5
1 – Disabled *
2 – N/A
3 – N/A
4 – Predefined with Titles
5 – Predefined w/o Titles
4) User passcode
Type value, press “ENT”
(max 16 numerical digits)
Default 1000