CALIBRATION

To restore defaults:

 

● Turn off the indicator. Then press HOLD and UNIT key simultaneously while the  

indicator is turning on (Do this while indicator is counting down)

 

● “

dEF 0

” will be displayed

 

● Press UNIT key to change to “

dEF 1

 

● Press HOLD key to confirm defaults restore

To display version number:

 

● Turn off the indicator. Then press UNIT and ON/OFF simultaneously while the  

indicator is turning on (Do this while indicator is counting down)

To Enter Calibration Settings

 

● Turn indicator off. Then press and hold the HOLD and UNIT key simultaneously 

while the indicator is turning on (Do this while indicator is counting down)

 

● The display will show [

CAL

 kg], meaning the unit is kg

 

● Press UNIT if you want to change the unit to lb; Display will change to [

CAL 

lb].

 

● Press HOLD to confirm your unit choice

 

● The display will most likely show [

FS 5

], meaning the max. capacity is 200kg, 

(see tables below for other variances)

 

● Press UNIT to change the capacity range to your scales capacity 

Calibration chart for kilogram usage:

Display

Max. Capacity

Division

Calibration Loading Weight

FS 0

15.00kg

0.005kg

10kg

FS 1

30.00kg

0.01kg

25kg

FS 2

60.00kg

0.02kg

50kg

FS 3

100.00kg

0.02kg

75kg

FS 4

150.00kg

0.05kg

100kg

FS 5

200.00kg

0.05kg

150kg

FS 6

custom

custom

custom

Calibration chart for pound usage:

Display

Max. Capacity

Division

Calibration Loading Weight

FS 0

33.00lb

0.01lb

22.05lb

FS 1

66.00lb

0.02lb

55.12lb

FS 2

132.00lb

0.05lb

110.25lb

FS 3

220.00lb

0.05lb

165.35lb

FS 4

330.0lb

0.1lb

220.5lb

FS 5

440.0lb

0.1lb

330.7lb

FS 6

custom

custom

custom

Содержание T902

Страница 1: ...www tufner com T902 INDICATOR USER S MANUAL...

Страница 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Страница 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Страница 6: ...table the indicator will Tare the scale and go into Net weighing mode when ZERO is pressed Hold There are three different hold functions you can choose the best option for you and set it the parameter...

Страница 7: ...unit is kg Press UNIT if you want to change the unit to lb Display will change to CAL lb Press HOLD to confirm your unit choice The display will most likely show FS 5 meaning the max capacity is 200kg...

Страница 8: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Страница 9: ...ight 2 Manual hold Grabs the current weight 3 Auto hold Automatically holds data when stable Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60...

Страница 10: ...rred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Soluti...

Страница 11: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Страница 12: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Страница 13: ...ECIFICATIONS CONTACT US Please e mail sales tufner com for any sales related questions Please e mail support tufner com for any support related questions Don t forget to visit our website at www tufne...

Отзывы: