5
CALIBRATION
To restore defaults:
● Turn off the indicator. Then press HOLD and UNIT key simultaneously while the
indicator is turning on (Do this while indicator is counting down)
● “
dEF 0
” will be displayed
● Press UNIT key to change to “
dEF 1
”
● Press HOLD key to confirm defaults restore
To display version number:
● Turn off the indicator. Then press UNIT and ON/OFF simultaneously while the
indicator is turning on (Do this while indicator is counting down)
To Enter Calibration Settings
● Turn indicator off. Then press and hold the HOLD and UNIT key simultaneously
while the indicator is turning on (Do this while indicator is counting down)
● The display will show [
CAL
kg], meaning the unit is kg
● Press UNIT if you want to change the unit to lb; Display will change to [
CAL
lb].
● Press HOLD to confirm your unit choice
● The display will most likely show [
FS 5
], meaning the max. capacity is 200kg,
(see tables below for other variances)
● Press UNIT to change the capacity range to your scales capacity
Calibration chart for kilogram usage:
Display
Max. Capacity
Division
Calibration Loading Weight
FS 0
15.00kg
0.005kg
10kg
FS 1
30.00kg
0.01kg
25kg
FS 2
60.00kg
0.02kg
50kg
FS 3
100.00kg
0.02kg
75kg
FS 4
150.00kg
0.05kg
100kg
FS 5
200.00kg
0.05kg
150kg
FS 6
custom
custom
custom
Calibration chart for pound usage:
Display
Max. Capacity
Division
Calibration Loading Weight
FS 0
33.00lb
0.01lb
22.05lb
FS 1
66.00lb
0.02lb
55.12lb
FS 2
132.00lb
0.05lb
110.25lb
FS 3
220.00lb
0.05lb
165.35lb
FS 4
330.0lb
0.1lb
220.5lb
FS 5
440.0lb
0.1lb
330.7lb
FS 6
custom
custom
custom