OPERATING INSTRUCTIONS

Powering On or Off

 

● Press ON/OFF key for 1 second to power on. Machine will self inspect and count 

from 

1 - 9

 before entering weighing mode and displaying 0

 

● Press ON/OFF key for 1 second to power off

 

● Indicator will auto power off when it has low battery

 

● Auto power-off in 10/20/60 minutes with no operation can be set in the parameters

Zero/Tare

 

● The zero function is used only when the scale is empty and is not at gross zero due 

to material build up

 

● If the weight on the scale is within the zero range (2% of the max. capacity) and 

stable and ZERO is pressed, the indicator will zero the scale  and stay in gross 
weighing mode

 

● The Tare function is used when you only wish to see the current change in weight, 

not the entire amount of weight that is on the scale

 

● If the scale is out of the zero range (above 2% of the max. capacity) and  

stable  the indicator will Tare the scale and go into Net weighing mode when ZERO 
is pressed

Hold

There are three different hold functions, you can choose the best option for you and set 
it the parameter settings

1. Peak Hold: Grabs the highest weight 

(for materials testing, ie. tension and pulling 

force)

 

● Press the HOLD key then add weight to the scale

 

● The indicator will show the highest weight it recorded and hold it on the screen until 

a higher weight is placed on the scale

2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate 

 

● While weighing, press HOLD and the indicator will hold the current weight on the 

screen until HOLD is pressed again

3. Auto Hold: If the weight on the scale is above 20d and is stable, the indicator will 

hold that weight on the screen for 3 seconds then go back to general weighing 

 

● Pressing the hold key is unnecessary, holding is done automatically when the scale 

is stable

Print

 

● To print press and hold the UNITS key when weighing is stable

 

● Communication Setting must be set to #2 for Print mode

Содержание T902

Страница 1: ...www tufner com T902 INDICATOR USER S MANUAL...

Страница 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Страница 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Страница 6: ...table the indicator will Tare the scale and go into Net weighing mode when ZERO is pressed Hold There are three different hold functions you can choose the best option for you and set it the parameter...

Страница 7: ...unit is kg Press UNIT if you want to change the unit to lb Display will change to CAL lb Press HOLD to confirm your unit choice The display will most likely show FS 5 meaning the max capacity is 200kg...

Страница 8: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Страница 9: ...ight 2 Manual hold Grabs the current weight 3 Auto hold Automatically holds data when stable Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60...

Страница 10: ...rred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Soluti...

Страница 11: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Страница 12: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Страница 13: ...ECIFICATIONS CONTACT US Please e mail sales tufner com for any sales related questions Please e mail support tufner com for any support related questions Don t forget to visit our website at www tufne...

Отзывы: