Remove and Clean Front Door:
(1) Push left pivot upward and inward.
(2) Push right pivot upward and inward.
Left Pivot
Front Door
Front Pin
(3) Pull front pin in the direction of arrow shown in the
photo.
(4) Remove front door.
The front door can be cleaned with 30 ° ± 5 ° C tap water or 60% to
70% alcohol, non-high-temperature cleaning.
Nut
15
■
Cleaning Couplings :
Loosen the nut and take the coupling off.
Clean it with ethanol 60 to 70 vol%
■
Note
(1) Remove it carefully without damaging the o-ring.
(2) Couplings cannot be autoclaved.
■
Note
Do not connect handpiece to the coupling when
o-ring is not properly fitted in its groove. Possible
damages to the o-ring and stuck in the coupling
may make it difficult for disassembling handpiece.
An o-ring could slip out of its groove.
Make sure all o-rings are properly seated in their
grooves before attaching a handpiece.
NG
OK
Upward
Blister
• The blister can be cleaned with 30˚
C±5
˚C tap water or 60% to 70% alcohol,
non-high-temperature cleaning.
• The blister can be used repeatedly, but should be changed once broken.
WARNING
■
Remove and Clean Blister
Remove it upward.
Tray
■
Cleaning the Tray:
(1) Take the tray out in the direction of arrow shown in the photo.
(2) Clean and replace new absorption pad.
• The tray cannot be autoclaved.
• Autoclaving will cause it to rust.
• Do not use disinfectants that contain
chlorine-based cleaners or benzalkonium
chloride or chlorinated aromatics.
• If this tray is not cleaned, oil will overflow
from the tray.
CAUTION
• Dispose of used absorption pad according to
regulations for medical waste.
• Push the tray all the way in, otherwise, oil
may leak and oil mist might be dispersed.
NOTE