• Explosion Hazard. Do not use near open flames or other ignition sources.
• Health Hazard. The area must be well ventilated. Breathing fumes could damage your health.
• The equipment must be properly grounded.
• Make sure the ground is properly connected.
• To prevent the spread of grave, life-threatening diseases like HIV and hepatitis B, autoclave
handpieces after performing regular maintenance using the TTcare 100+.
• Avoid the risks of electrical shock, equipment damage and fire during an electrical storm: Turn the
TTcare 100+ off and do not touch it or its cord.
• Wear surgical gloves, mask and protective eyewear to operate and clean the TTcare 100+.
WARNINGS
Electromagnetic wave interference caused by cellular phones, transceivers, remote controls and
similar transmission devices could cause the equipment to operate randomly. All devices which
transmit electromagnetic waves located near the work area should be turned off.
CAUTIONS
• Set the TTcare 100+ on a level and stable surface.
• Depending on the type of handpiece and coupling, there could be a release of oil mist. In this
case, lower the pressure to 4 bar.
• Oil will leak or spray out if the cans are not properly installed.
• Make sure the head is flat against the nozzle guide. If it is tilted, the nozzle might be bent or oil
might spray out.
• Oil will spray out of the chuck nozzle if the handpiece is removed too soon (before the beep).
NOTES
ATTENTION
TTBIO CORP. will not be responsible for accidents, instrument damage, or
(1) repairs made by personnel not authorized by TTBIO CORP.
(2) any changes, modifications, or alterations of its products.
(3) maintenance or repairs using parts or components other than those specified by TTBIO CORP. and
other than in their original condition
(4) operating the instrument in ways other than the operating procedures described in this manual or
resulting from the safety precautions and warnings in this manual not being observed
(5) workplace conditions and environment or installation conditions which do not conform to those
stated in this manual such as improper electrical power supply
(6) fires, earthquakes, floods, lightning, natural disasters, or acts of God.
1. Warnings, Cautions and Notes
• For some handpieces, an oil mist may be released if a burr is not inserted.
• Do not damage the o-ring on the coupling when attaching a handpiece. This could result in oil
spraying out or poor performance.
• Do not select a coupling that does not have a handpiece attached to it. This would result in oil
spraying out when the Start Switch is pressed.
• If the unit will not be used for a while, unplug it and close the main air valve.
• Please insure you use the correct adaptor for the instrument.
• Do not use couplings made for maintenance equipment other than the TTcare 100+. This could
result in release of oil mist or poor handpiece maintenance.
• Dispose of used oil pads according to regulations for medical waste.
• Do not fail to push the tray all the way in. Otherwise, oil could leak out or oil mist might be
released.
2. Specifications
Name
TTcare 100+
AC 100-240 V 50/60 Hz
60VA max.
250V 2A Slow Blow Type
4~6 bar
4 bar
40 NL/ min
Approx. 9 kg
Width 295 × Height 385 × Depth 295 mm
Rating
Power Consumption
Fuse
Air Pressure
Recommended Air Pressure
Air Flow Rate
Weight
Size (including regulator)
Protection category (Dust-proof;
water-proof)
IP20
Operating environment
Permitted in interior rooms
15 to 40°C (59 to 104°F)
25 to 90 %
2000 m (6560 feet)
Ambient temperature
Temperature
Humidity
Max. altitude
Transportation and storage conditions
-20 to 70°C (-4 to 158°F)
5 to 95 %
700 to 1060 hPa
Temperature
Humidity
Atmospheric pressure