background image

  

Sensors  

     

Natural Wet Bulb Thermometer 

 

Sensors 

Natural Wet Bulb Thermometer 

The natural wet bulb thermometer 
gives an indication of the effects of 
humidity on an individual. Relative 
humidity and wind speed are taken 
into account by measuring the 
amount of evaporative cooling taking 
place at a thermometer covered with 
a moistened wick. 

A        B        C 

A

. Globe thermometer 

B

. Natural wet bulb 

thermometer 

C

. Dry bulb thermometer 

 

Check Wick and Fill Natural Wet Bulb 

The QUESTemp° Heat Stress Monitor uses a cotton wick immersed into a reservoir containing 
distilled water. 

DO NOT

 use ordinary tap water, as the contaminants that are left behind after 

evaporation will shorten the life of the wick and cause high readings. If the wick is discolored it 
should be replaced. To 

replace the wick

, slide the old wick off the top of the sensor. Place a 

new wick over the sensor, making sure that the bottom of the wick is down in the reservoir.  
 

 

Wick 

Reservoir cover 

Reservoir 

Filling wet bulb reservoir 

1.

 

Remove reservoir cover and fill with 

distilled or de-ionized water. 

2.

 

Replace reservoir cover. 

Figure 1-4:  Filling wet bulb reservoir 

Globe Thermometer 

The globe thermometer (left position) gives an indication of the radiant heat exposure on an 
individual due to either direct sunlight or hot objects in the environment. This is accomplished by 
placing a temperature sensor inside a blackened copper sphere and measuring the temperature 
rise. The WBGT index is based on the response of a 6 inch diameter globe. The QUESTemp° 
uses a 2-inch diameter globe for a faster response time. The temperature of the 2-inch globe is 
correlated to match that of a 6 inch globe. 

Sensors

 

Содержание QUESTEMP 34

Страница 1: ......

Страница 2: ......

Страница 3: ...obe 19 Operational Check 20 Power options 20 9 Volt Alkaline Battery Replacement 20 Approved 9 Volt Batteries 21 NiMH Battery Pack 21 Appendix A Specifications 22 Measurements 22 Data Logging 22 Langu...

Страница 4: ...en 13 1 13 Air Flow screen 14 1 14 Stay times ACGIH 14 1 15 Time Date screen 14 1 16 Battery Memory screen 14 1 17 Navy Stay time screen 15 1 18 EPRI Stay time screen 15 1 19 Sample printout page 1 16...

Страница 5: ...els support RS 232 communications and works with the optional software analysis program TSI Detection Management Software DMS With the QUESTemp 36 you have the capability to display stay times in orde...

Страница 6: ...ground 3 Power On the QUESTemp 34 36 and check the battery If the battery voltage displayed during the power on sequence is less than or equal to 6 4 volts replace or recharge the batteries 4 Be aware...

Страница 7: ...splay to view the next available sensor bar indicated in the upper right corner of the display Up Arrow key Changes items appearing in the display Scrolls up Down Arrow key Changes items appearing in...

Страница 8: ...ive language select Setup from the main menu Press the Down arrow repeatedly until English or the appropriate language appears Then repeatedly press the I O enter key to toggle through the languages O...

Страница 9: ...discolored it should be replaced To replace the wick slide the old wick off the top of the sensor Place a new wick over the sensor making sure that the bottom of the wick is down in the reservoir Wic...

Страница 10: ...is used in the outdoor WBGT calculation when a high solar radiant heat load may be present The series of white plates surrounding the sensor shield it from radiant heat Relative Humidity Sensor A rel...

Страница 11: ...ar by pressing and releasing the enter key The upper right corner of the display shows the current sensor bar 1 refers to the top sensor bar 2 and 3 are labeled on the side of the unit W indicates the...

Страница 12: ...etermine the thermal comfort indices Predicted Mean Vote PMV and Predicted Percent Dissatisfied PPD with the QUESTemp 36 model only Wet Bulb Globe Temperature The WBGT is a weighted average of the thr...

Страница 13: ...e Air Probe uses an omni directional anemometer sensor that measures air flow between 0 and 20 meters per second in 0 1 m s increments See Air Flow Functionality for more details Thermal Comfort QUEST...

Страница 14: ...R key while a 3 2 1 countdown is shown in the lower right corner of the display Then the menu screen will appear see Figure 1 3 for an example Setup The setup screen is used to change the following pa...

Страница 15: ...ng the Up Down Arrows and I O Enter keys to modify each number NOTE At any time you can move back one level by pressing the Run Stop key 4 Exit Setup by pressing the Run Stop key Print QT 34 36 allows...

Страница 16: ...Ti 66 7 F 1 WBGTo 68 6 F Run indicator Figure 1 8 Run mode indicator 2 End the session by pressing the Run Stop key again The session will stop recording when the asterisk is no longer displayed NOTE...

Страница 17: ...layed and all three sensor bars are attached An asterisk in the lower right corner indicates that the unit is in the run mode and is logging data The following measurements can be accessed on the disp...

Страница 18: ...y Figure 1 16 Battery Memory screen NOTE A series of dashes appear in the display if one of the following occur The Heat Index or Humidex is outside of its allowable range The temperature is outside o...

Страница 19: ...below NOTE 4 01 indicates greater than four hours L M H 1 4 01 3 00 1 30 Light Medium and Heavy maximum hours displayed Figure 1 18 EPRI Stay time screen Data Logging Data from each sensor is recorded...

Страница 20: ...___________ Facility _________________________ Session 3 Start 21 FEB 08 11 07 32 Department _________________________ Stop 21 FEB 08 11 10 15 Job _________________________ Printed 21 FEB 08 11 16 00...

Страница 21: ...WBGTi WBGTo WET DRY GLOBE RH HI L M H VH 11 08 79 9 79 3 69 4 98 9 104 5 15 0 60 45 30 15 11 09 80 8 80 2 70 3 99 2 105 6 15 0 60 45 30 15 Session 3 Page 4 Sensor 3 Degrees Fahrenheit Stay Times ACGI...

Страница 22: ...e be cautious if measuring in ducts The Air Probe may be either hand held or mounted behind the QUESTemp 36 using the mounting bracket hooked to the sensor bar beneath the center bulb sensor See Figur...

Страница 23: ...he Sensor 1 data therefore air flow data will only be reported if a sensor bar is plugged in Air flow is recorded during the run mode at the interval the QUESTemp 36 is setup for Batteries for Air Pro...

Страница 24: ...ower supply method is changed The up position is for the 9 volt battery The down position allows for either the AC adapter or the rechargeable batteries The AC adapter will trickle charge the recharge...

Страница 25: ...le battery pack is charged in the instrument using TSI s AC 120V AC to 9V DC adapter part 015 910 or 220V AC to 9V DC adapter part 015 680 A discharged battery pack requires an overnight charge of 16...

Страница 26: ...talian German Housing Designed water resistant to a light rain or mist If rain is frequent best practice would be to remote the sensor bar and keep the instrument sheltered Size Height 9 2 in 23 5 cm...

Страница 27: ...line 140 hours Rechargeable Nickel Metal Hydride 300 hours Adding additional sensor bars reduces battery life Charge Time NiMH Battery Pack 16 hours charge in the unit Air Probe Accessory QUESTemp 36...

Страница 28: ...hargeable battery Type Manufacturer U9V Ultralife MN1604 Duracell 522 or EN22 or 6LR61 Energizer A1604 or BR232 Rayovac 6LR61 or 6AM6 Panasonic Rechargeable battery Integral NiMH battery pack type DC2...

Страница 29: ...Moderate Heavy Very Heavy 75 to 100 31 0 28 0 26 0 23 5 50 to 75 31 0 29 0 27 5 25 5 25 to 50 32 0 30 0 29 0 28 0 0 to 25 32 5 31 5 30 5 30 0 Work and recovery Action Limit Light Moderate Heavy Very H...

Страница 30: ...l Hygienists Threshold Limit Values and Biological Exposure Indices for 2008 Reprinted with permission United States Navy Physiological Heat Exposure Limits PHEL Time Table Without the presence of fue...

Страница 31: ...0 1 35 1 20 1 05 0 50 0 35 105 0 1 40 1 30 1 15 1 00 0 45 0 30 106 0 1 35 1 25 1 10 0 55 0 45 0 30 107 0 1 30 1 15 1 05 0 50 0 40 0 25 108 0 1 20 1 10 1 00 0 50 0 35 0 25 109 0 1 15 1 05 0 55 0 45 0 3...

Страница 32: ...s A time of 4 01 indicates greater than 4 hours WBGT C Light Moderate Heavy 28 4 01 4 01 3 00 29 4 01 4 00 2 00 30 4 01 3 00 1 30 31 4 01 2 00 1 15 32 4 00 1 30 1 00 33 3 30 1 15 0 45 34 3 00 1 00 0 4...

Страница 33: ...ot shielded remote sensor cable 053 926 200 foot shielded remote sensor cable 053 927 Serial computer cable 054 715 Parallel printer cable 056 875 120 VAC to 9 VDC adapter 015 910 220 VAC to 9 VDC ada...

Страница 34: ...up parameters Refer to the DMS User Manual for further instructions Downloading your sessions in DMS Once the QUESTemp 34 36 is communicating and run stops are stored on the instrument this data may b...

Страница 35: ...and then select the Model Type by clicking on QT 34 or QT 36 See and You are now ready to setup or download your instrument 5 Click in the Data Finder checkbox see and then select button see NOTE When...

Страница 36: ...ude analyzing and printing a quick report See below for more details Analyze data 1 In the Data finder page click a heat stress session by expanding the Family heat stress and selecting the session NO...

Страница 37: ...he order of the charts tables which appear in the panel layout Also when a chart table is selected in the arrange panels palette the associated data panel is selected The resize handles are applied an...

Страница 38: ...quence in which they appear on the screen Printing reports while in the report view page you have two options outlined below Click on the icon from the title bar Click on the icon from the session rep...

Страница 39: ...tsi com Europe Middle East and Africa Telephone 49 241 52303 0 Fax 49 241 52303 49 E mail tsigmbh tsi com Service Contact Information If your instrument does not operate properly or if you are returni...

Страница 40: ...lling address A shipping address Use the original packing material to return the instrument to TSI If you no longer have the original packing material seal off any ports to prevent debris from enterin...

Страница 41: ...any person other than Seller The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR...

Страница 42: ...more information USA Tel 1 800 874 2811 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 0...

Отзывы: