9
Chapter 4
Maintenance
The Model 5725 requires very little maintenance to keep it performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 5725 to TSI for annual recalibration.
Please contact one of TSI’s offices or your local distributor to make service
arrangements and to receive a Return Material Authorization (RMA)
number. To fill out an online RMA form, visit TSI’s website at
U.S. & International
TSI Incorporated
500 Cardigan Road
Shoreview MN 55126-3996
Tel: (800) 874-2811
(651)
490-2811
Fax: (651) 490-3824
The Model 5725 can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make minor
changes in calibration to match a user's calibration standards. The field
adjustment is NOT intended as a complete calibration capability. For
complete, multiple-point calibration and certification, the instrument must be
returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with a soft
cloth and isopropyl alcohol or a mild detergent. Never immerse the Model
5725. If the enclosure of the Model 5725 or the AC adapter becomes broken,
it must be replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month to
prevent damage due to battery leakage.