Using Serial Data Commands
C-17
Example
Response
RF
00AC
00AC can be converted to binary: 0000 0000 1010 1100
This indicates that:
Internal temperature is greater than 40°C
Autocal failed
Excessive sample concentration
Sheath flow out of range
RI
—Read Input From Pins 1, 2, 3 and 7, 8 of the
I/O Connector
Pins 1, 2, and 3 are digital inputs and are reported as a binary number 000
to 111, where a 1 indicates 5 VDC level on the pin. (Digital ground pins for
the inputs are 9 and 10.) Pins 7 and 8 are analog inputs referenced to pin
15 analog ground. The analog values are reported as a decimal number
from 0.0 to 5.000 VDC. All readings are instantaneous.
Example
Response
RI
100,2.43,1.93
Pin 1 has logic level high (5 VDC). Pins 2 and 3 are logic level low (0
VDC). Pin 7 is 2.43 VDC. Pin 8 is 1.93 VDC.
RL
—Read Laser Power
Reads the current laser power output in percentage (%) of maximum
milliwatts and current in milliamps (ma)
Example
Response
RL
75.0, 65.3
Indicates the laser power output is 75 percent of maximum output and
current is 65.3 ma.
RO
—Read Accumulated On Time
Reads the accumulated on-time of the instrument (in hours). The time is
updated once an hour and stored in EEROM.
Example
Response
RO
72
Indicates the Model 3321 has been on for approximately 72 hours.
Содержание SIZER 3321
Страница 1: ...AERODYNAMIC PARTICLE SIZER SPECTROMETER MODEL 3321 USER S MANUAL P N 1930092 REVISION H SEPTEMBER 2013...
Страница 2: ......
Страница 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 98: ......
Страница 100: ......