3-4
Model 3321 Aerodynamic Particle Sizer
Spectrometer
Inlet Nozzle
The inlet nozzle on the top of the Model 3321 is designed so that aerosol
can be sampled from a chamber or open air with good efficiency. Tubing
can be attached to the inlet to sample when necessary. The inlet is
.746 inches (18.9 mm) in diameter for use with ¾-inch Swagelok
®
-type
connectors or with slightly smaller inner diameter flexible tubing.
Note
: Conductive tubing is recommended for use with the APS
™
spectrometer to minimize particle loss due to electrostatic charge.
Suitable tubing is available from TSI.
Indicators
There are five status LEDs on the APS
™
spectrometer: PARTICLE, HI
CONC, FLOW, LASER, and POWER.
The amber PARTICLE LED blinks once each time a particle passes
through the sensor. In normal room air, the LED will appear to be lit
constantly. When sampling aerosols in low concentrations, the LED will
appear to flicker.
The amber HI CONC LED indicates that the concentration of particles
being sampled is above the recommended level to prevent coincidence
(see Chapte
r 5, “
”). When this LED is lit, many of
the particles are ignored in the counting process since they cannot be
accurately sized. The default setting is 1000 particles/cm
3
.
The green FLOW LED indicates that both the sample and sheath
airflows are within their specified range.
The green LASER LED indicates that the laser is on and functioning
properly. A flashing LED may mean that the laser is not able to operate
at the set power. See
command.
The green POWER LED indicates that power is supplied to the
instrument.
®
Swagelok is a registered trademark of Swagelok® Companies, Solon, Ohio.
Содержание SIZER 3321
Страница 1: ...AERODYNAMIC PARTICLE SIZER SPECTROMETER MODEL 3321 USER S MANUAL P N 1930092 REVISION H SEPTEMBER 2013...
Страница 2: ......
Страница 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 98: ......
Страница 100: ......