Maintenance
A-5
14.
The finish on the outside of the nozzle should be flat black. Remove
any coating on the outer nozzle with a clean, lint-free, soft cloth and
clean water.
15.
Check that the nozzle is clear by holding it up to a light and checking
for a clear view through the nozzle orifice.
16.
It is not recommended that you clean the inside of the outer nozzle
with other than clean compressed air. Using a swab to clean may
result in leaving difficult-to-remove fibers inside the nozzle or
damaging the inner nozzle centering ring.
17.
Reinsert the nozzle block into the optics block. The nozzle block has
a slot that aligns with a pin in the optics block. Tighten the retaining
ring so that the nozzle block can still rotate freely. Rotate the nozzle
until the pin catches the slot in the nozzle. Tighten the retaining ring
until the nozzle block is firmly seated.
18.
Reinsert the inner nozzle and make sure that it is seated firmly in the
nozzle housing.
19.
Reattach the tubing to the nozzle mount, noting the labels on each
tube.
20.
Tilt the digital printed circuit board back down until the ribbon cable
connector can be plugged back into the board. Plug in the ribbon
cable connector.
21.
Tilt the digital printed circuit board down fully and fasten with the
screws provided.
22.
Replace the outer inlet in the nozzle mount. Note that there is an
alignment pin that matches a slot in the side of the outer inlet. Rotate
the outer inlet until the inlet seats on the alignment pin.
23.
Replace the cover on the instrument.
24.
Tighten the six screws holding the cover in place.
25.
Replace the outer inlet retaining ring and tighten firmly.
26.
Insert the power cord into the power entry connector on the back of
the instrument.
27.
Apply power to the instrument with the switch on the back panel.
28.
Check that the flows are correct.
29.
Generate a known size of monodisperse aerosol (such as
Polystyrene latex
–PSL) and make sure that the calibration has not
been altered by the cleaning procedure.
30.
If the instrument does not size correctly, check the parts you have
cleaned and make sure that they are assembled correctly: alignment
pins in slots and retaining rings holding parts seated firmly.
31.
If the instrument still does not size correctly, it may have to be
recalibrated.
Содержание SIZER 3321
Страница 1: ...AERODYNAMIC PARTICLE SIZER SPECTROMETER MODEL 3321 USER S MANUAL P N 1930092 REVISION H SEPTEMBER 2013...
Страница 2: ......
Страница 10: ...viii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 14: ...xii Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 16: ...xiv Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 26: ...2 6 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 36: ...3 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 50: ...5 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 62: ...A 10 Model 3321 Aerodynamic Particle Sizer Spectrometer This page intentionally left blank...
Страница 98: ......
Страница 100: ......