TSI Instruments PORTACOUNT 8020 Скачать руководство пользователя страница 5

P

ORTA

C

OUNT

®

 Plus Model 8020 Technical Addendum

revised 7-27-2001

TSI Incorporated, St. Paul, Minnesota, USA

5

Copyright © 2001 all rights reserved

DIP Switch 4: Memory Lock

Switch 4 allows you to lock the P

ORTA

C

OUNT

 Plus memory so that the internal settings for

the sampling times, fit factor pass level and number of exercises cannot be changed from
the P

ORTA

C

OUNT

 Plus Keypad or by using the MODIFY program software supplied with

the instrument.  The reason you might want to lock memory is to prevent unauthorized
persons from altering the fit test protocol.  The factory setting for switch 4 is ON which
means that memory lock feature is OFF and the internal settings can be changed using the
Keypad or software.  The FitPlus software is not affected by DIP switch 4.

Memory

Switch 4

Unlocked

ON

Locked

OFF

DIP Switch 5: Factory Checkout

Switch 5 is used by TSI when performing factory checkout of the P

ORTA

C

OUNT

 Plus and

should always be in the ON position for normal use.

Mode

Switch 5

Checkout

OFF

Normal

ON

DIP Switch 6: Reset Factory Settings on Power Up

Switch 6 should always be left ON unless you want to reset all the internal settings back
to the way they were when the P

ORTA

C

OUNT

 Plus was new.

When switch 6 is OFF, the P

ORTA

C

OUNT

 Plus will reset back to the factory settings when

the power is turned on.  Once the power is turned back on, you will be able to change the
internal settings (unless switch 4 is OFF), but they will be reset to the factory values the
next time the power is turned off and then on again.

If you want to reset the internal settings back to the factory values, move switch 6 to the
OFF position and then start the P

ORTA

C

OUNT

 Plus.  Now turn the P

ORTA

C

OUNT

 Plus off and

change switch 6 back to the ON position.  The instrument will be reset and you will be
able to permanently change the internal settings.  The factory settings are:

Fit Factor Pass Level:

100

Number of exercises:

8

Ambient Purge Time

4 sec

Ambient Sample Time

5 sec

Mask Purge Time

11 sec

Mask Sample Time, All exercises:

40 sec

DIP Switch 7: Reserved

Switch 7 must always be left ON.

DIP Switch 8: Use CTS

When switch 8 is on (the factory setting), the P

ORTA

C

OUNT

 Plus will not require a Clear

To Send (CTS) signal in order to send data through the RS-232C Data Port to a

Содержание PORTACOUNT 8020

Страница 1: ...del 8020 Technical Addendum revised 7 27 2001 TSI Incorporated St Paul Minnesota USA 1 Copyright 2001 all rights reserved www tsi com PORTACOUNT Plus Model 8020 Technical Addendum includes the N95 Com...

Страница 2: ...espirator fit testing Persons who will find this document of interest include respirator manufacturers respiratory protection researchers and persons conducting general ultrafine particle studies not...

Страница 3: ...p and bottom halves of the enclosure Follow the instructions below 1 Turn the PORTACOUNT Plus off 2 Turn the PORTACOUNT Plus over so that it is face down and locate the four screws on the bottom that...

Страница 4: ...Setting 1 ON 2 OFF 3 ON 4 ON 5 ON 6 ON 7 ON 8 ON The function of each of the DIP Switches is as follows Factory settings are shown in BOLD type DIP Switches 1 2 and 3 Baud Rate Switches 1 2 and 3 con...

Страница 5: ...actory Settings on Power Up Switch 6 should always be left ON unless you want to reset all the internal settings back to the way they were when the PORTACOUNT Plus was new When switch 6 is OFF the POR...

Страница 6: ...esota USA 6 Copyright 2001 all rights reserved peripheral device When switch 8 is OFF the CTS signal will be required The CTS signal is always required when the PORTACOUNT Plus is in External Control...

Страница 7: ...cc FF 3 505 PASS Fit factor exercise 3 Mask 4 10 cc Ambient 5100 cc FF 4 1231 PASS Fit factor exercise 4 Mask 7 90 cc Ambient 4900 cc FF 5 610 PASS Fit factor exercise 5 Mask 13 50 cc Ambient 4800 cc...

Страница 8: ...n that shows the internal memory settings The settings reflect how the PORTACOUNT Plus memory is set at the moment and can be changed For example when all settings are at the factory settings the data...

Страница 9: ...tches are considered ON if set to 1 and OFF if set to 0 DIP Switches are numbered 1 8 from left to right on the printout and 1 8 from top to bottom on the circuit board Output When a Battery Runs Low...

Страница 10: ...dular connector Contact 1 is on the left when viewed as shown Data Port Pin Assignments Contact Number Function Direction 1 Not used 2 Ready DTR From PORTACOUNT Plus 3 Receive data Rx To PORTACOUNT Pl...

Страница 11: ...of these cables are always female Serial interface connectors on the back of the computer are always male Pin 1 on the PORTACOUNT plus Data Port is on the left as you look at the Data Port connector T...

Страница 12: ...COUNT Plus will be terminated with both an ASCII carriage return cr and a line feed lf character When the PORTACOUNT Plus receives a command that is not understood or has values that are outside the a...

Страница 13: ...haracters sent are ignored until the first J cr is received Response from PORTACOUNT Plus OK cr lf When External Control Mode is invoked certain parameters are automatically set as follows The interna...

Страница 14: ...when External Control Mode is first invoked Response from PORTACOUNT Plus VO cr lf ZD cr Disable Continuous Data Transmission This command tells the PORTACOUNT Plus to stop sending continuous particle...

Страница 15: ...tery Bad Pulse Good RBB cr lf Battery Bad Pulse Bad S cr Request Settings This command forces the PORTACOUNT Plus to transmit all current settings for sample times fit factor pass level serial number...

Страница 16: ...vvv is the fit factor pass level contained in that memory location The maximum value allowed for the fit factor pass level is 64000 There is one space after the SP Serial Number The PORTACOUNT Plus se...

Страница 17: ...to set the mask sample time for exercise 4 to 40 seconds send the command PTM0440 cr The PORTACOUNT Plus will respond with an echo of the same command PTM0440 cr lf PTA00vv cr Set Ambient Sample Time...

Страница 18: ...t Factor Pass Level The PORTACOUNT Plus stores a set of 12 fit factor pass levels in memory The operator is allowed to scroll through and select one of these values as the current fit factor pass leve...

Страница 19: ...ACOUNT Plus This command clears the Display and then displays a fit factor pass level The FIT FACTOR PASS LEVEL message is turned ON xxxxxx is the fit factor pass level to display and can be in the ra...

Страница 20: ...echoing A000740 0 cr lf Nxx cr Display Exercise Number on PORTACOUNT Plus This command displays an exercise number on the PORTACOUNT Plus Display xx is the number to display and must be in the range f...

Страница 21: ...Turn SERVICE message OFF Bit 5 LOW PARTICLE Message 1 Turn LOW PARTICLE message ON 0 Turn LOW PARTICLE message OFF Bit 6 LOW BATTERY Message 1 Turn LOW BATTERY message ON 0 Turn LOW BATTERY message OF...

Страница 22: ...000001 cr The PORTACOUNT Plus responds I00000001 cr lf To turn the FIT FACTOR message and the PASS message ON at the same time bit 3 must be a 1 and bit 8 must be a 1 send I00100001 cr The PORTACOUNT...

Страница 23: ...com to obtain a copy To operate with the N95 Companion the PortaCount Plus must have the revision A modifications The tag on the bottom of the PortaCount Plus will show MODEL 8020A if it is N95 Compan...

Страница 24: ...ers Application Note ITI 041 PortaCount Theory of Operation Application Note ITI 048 N95 Companion Model 8095 Theory of Operation Application Note ITI 053 Penetration of Ambient Aerosols Through Respi...

Страница 25: ...rights reserved The valve inside the PortaCount is no longer used and is left un energized sample tube open The valve control is automatically re routed to the N 95 Companion valve located inside the...

Страница 26: ...ced by a factor of 10 Pass Fail setting for PortaCount alone Pass Fail setting with N95 Companion Comment 0 0 Pass Fail indicator OFF 1 to 10 1 Minimum pass fail 1 50 5 100 10 200 20 500 50 1000 100 2...

Страница 27: ...to ground to signal PortaCount that N95 Companion is attached 4 5 VDC Solenoid valve voltage feed from PortaCount High current voltage supply for solenoid valve in pendant 5 Solenoid valve control fro...

Страница 28: ...PORTACOUNT Plus Model 8020 Technical Addendum revised 7 27 2001 TSI Incorporated St Paul Minnesota USA 28 Copyright 2001 all rights reserved...

Отзывы: