Plus Model 8020 Technical Addendum
revised 7-27-2001
TSI Incorporated, St. Paul, Minnesota, USA
24
Copyright © 2001 all rights reserved
The N95-Companion is used as a particle conditioner in between the respirator and the
PortaCount. The 0.04 micrometer particles that are allowed to pass through to the
PortaCount are in the size range that is effectively stopped by low efficiency filter media
as well as high efficiency media. By only allowing the PortaCount to "see" particles that
can only enter the respirator via a face seal leak, the N95-Companion makes low
efficiency media look like high efficiency media to the PortaCount. Since both the
ambient and respirator sample pass through the N95-Companion, the ratio of outside
concentration vs. inside concentration (fit factor) is valid.
For more information visit the
and review the following documents:
Mechanisms of Filtration for High Efficiency Fibrous Filters
(Application Note ITI-041)
PortaCount Theory of Operation
(Application Note ITI-048)
N95-Companion Model 8095 Theory of Operation
(Application Note ITI-053)
Penetration of Ambient Aerosols Through Respirator Faceseal Leaks
(Application Note ITI-055)
Fit Testing Using Size-Selected Aerosol
(Application Note ITI-062)
PortaCount Internal Program Configuration
When the PortaCount Plus senses the presence of the N95-Companion it automatically
resets itself and runs with special configuration parameters designed to function with the
N95-Companion. The changes include:
•
Sample timing changes to accommodate the longer N95-Companion sampling
requirements.
Ambient purge = 6 seconds
Ambient sample = 15 sec
Mask purge = 15 sec
Mask sample 50 sec,
•
The maximum fit factor that can be displayed is 200. Any fit factor measured to be
above 200 is displayed as 200 with the words "Fit Factor" flashing. The 200 limit
was necessary because the particle concentrations output by the N95-Companion are
too low to allow accurate fit factor measurements at higher levels.
•
The pump inside the N95-Companion runs in addition to the pump inside the
PortaCount.