17. Attach ferrite to 

cable with one wrap 
of the cable and 
snap the ferrite 
together. 

 

18. Attach heated inlet 

cable to heated inlet 
control module (J4).  

 

19. If using a Thiamis 1000 Node, route the other side of 

the cable to the power input on the Thiamis node 
(green connector). If you are not using the Thiamis 
1000 telemetry node, this cable can remain 
disconnected. 

 

Heated Inlet Removal 

1.  Unscrew the top 

knurled ring/collar 
from the lower 
mounting collar. 

 

2.  Inside the 

Enclosure, 
disconnect the 
Heated Inlet cable 
(note that the 
connector has a 
lock-release on it). 

 

3.  Remove the Ferrite 

Clamp Assembly. 
Note that the clamp 
is released by 
inserting and 
twisting a small flat 
blade screwdriver. 
Also, for purposes 
of re-assembly, 
note that the cable 
twists two times 
through the Ferrite. 

 

4.  Carefully pull the 

Heated Inlet 
Assembly up. 

 

5.  Carefully pull the 

Heated Inlet cable 
and connector 
through the slot in 
the Mounting Ring.  
 
Now the Heated 
Inlet Assembly can 
then be completely 
removed. 

 

Содержание DustTrak 854041

Страница 1: ...___ TSI and TSI logo are registered trademarks of TSI Incorporated DustTrak is a trademark of TSI Incorporated HEATED INLET SAMPLE CONDITIONER ACCESSORY 854041 FOR DUSTTRAK ENVIRONMENTAL MONITORS INSTALLATION GUIDE ...

Страница 2: ......

Страница 3: ...owing table are present 2 Contact TSI immediately if items are missing or broken 3 Additional items may be included if you ordered accessories or spare parts Qty Item Description Reference Picture 1 Heated Inlet Column 1 RH Temp Sensor 1 RH Temp Sensor Gland 1 Heated Inlet Control Module 1 Omni Direction Inlet Qty Item Description Reference Picture 1 Water Trap 1 Power and Communication Cable 1 Co...

Страница 4: ... ensure a waterproof seal 3 Insert RH Temp Sensor through gland and tighten gland nut Extend probe 2 75 7 cm beyond nut 4 Attach the sun shield to the top of the RH Temp Sensor as shown Sun Shield should not touch sintered cap of RH Temp Probe 5 Attach RH Temp Sensor cable to heated inlet control module J3 6 Attach the two larger gray connectors to side of the photometer 7 Attach the two smaller w...

Страница 5: ...umn 12 Attach the inlet to the column 13 Install O ring and route cable through the mounting ring Pull excess cable through to prevent pinching of cable when inlet column is installed 14 Set inlet column on top of photometer mounting ring rotate slowly until alignment pin mates with the hole in the bottom of heated inlet column 15 Slide column retainer ring down heated inlet column 16 Hand tighten...

Страница 6: ...nscrew the top knurled ring collar from the lower mounting collar 2 Inside the Enclosure disconnect the Heated Inlet cable note that the connector has a lock release on it 3 Remove the Ferrite Clamp Assembly Note that the clamp is released by inserting and twisting a small flat blade screwdriver Also for purposes of re assembly note that the cable twists two times through the Ferrite 4 Carefully p...

Страница 7: ...escribes the status of the heated inlet Solid Green Inlet temp is 1 C below set point and controlling to maintain set point Blinking Green Inlet temp is between 1 to 5 C below set point and controlling to improve Blinking Red Inlet temp is more than 5 C below set point and controlling to improve This will occur when unit is first turned on and coming to temperature Solid Red Sensor unplugged or ha...

Страница 8: ... more information USA Tel 1 800 680 1220 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 6008577 Rev D 2018 TSI Incorporated Printed in U S A ...

Отзывы: