Installing the Heated Inlet
(854041)
1. Attach heated inlet control module to the front of the
Photometer. This can be done using the 2x 6-32 x
5/16 Socket Head Cap Screws and 2x flat washers.
Use a 7/64” ball driver.
2. Push out elastic plug and install RH/Temp Sensor
sealing gland. Removable nut should be installed on
the inside of the enclosure.
W A R N I N G
Water gasket on gland must form a good
seal with the exterior of the enclosure to
ensure a waterproof seal.
3. Insert RH/Temp
Sensor through
gland and tighten
gland nut. Extend
probe ~2.75” (7 cm)
beyond nut.
4. Attach the sun
shield to the top of
the RH/Temp
Sensor as shown.
Sun Shield should
not touch sintered
cap of RH/Temp
Probe.
5. Attach RH/Temp
Sensor cable to
heated inlet control
module (J3).
6. Attach the two
larger (gray)
connectors to side
of the photometer.
7. Attach the two
smaller (white)
connectors to the
Power
/
Communication
inputs (J1 and J2)
of the heated inlet
control module.
8. Install the water trap bottle onto the side of the inlet
as shown.
NOTE
For best results, use PTFE Thread Sealant
Tape on the water bottle threads.