AXD530 MicroManometer Specifications
Measurement Units
Pa
Pascals
in. H2O
inches of Water Column
fpm
feet
per
minute
m/s
meters
per
second
Resolution
Pressure
Measurement
0.01
in.
H2O (-1.00 to +10.00)
1 Pa
Display resolutions for Velocity units are:
0.01
(0
to
99.99)
0.1
(100
to
999.9)
1
(1000
to
9,999)
scientific
notation (10,000+)
Range
Pressure Measurement
-1.00 to +10.00 in. H2O
-250 to 2500 Pascals
Velocity Measurement
400 to 12,664 fpm
([
]
displayed)
1.29 to 64.5 m/s
Operating Temperature
14°F to 122°F (-10°C to 50°C)
Storage Temperature
-40°F to 176°F (-40°C to 80°C)
Overpressure Limit
20 psi or 137 kPa or
553
in.
H
2
O maximum
Accuracy
(Factory Calibration conditions)
For pressure measurement only,
after
zeroing.
±(1% of indicated r2 x
resolution)
Display
.45 inch high, 4 digit, 7 segment LCD, no
backlight
Low Battery Indicator
4
Units
Annunciators
Physical
Dimensions
7.6 x 3.2 x 1.3 inches
193 x 81 x 33 millimeters
Weight
11.6 oz. (330g) with batteries
Power Source
3 AA-size alkaline batteries
Battery Life
15 hours with continuous use
Model
Part
No.
AXD530
AXD530 meter,
632-530-110
protective pouch,
3 AA-size batteries,
operator’s manual,
calibration data sheet,
and two-year limited
warranty.
AXD532
Manometer kit with
632-613-001
AXD530 meter, 18
inch pitot probe, carrying
case, 3 AA-size batteries,
manual, 2 static pressure
taps, 100 duct plugs and
two-year limited warranty.
Pitot Probes
12 inch pitot probe
634-634-000
18 inch pitot probe
634-634-001
24 inch pitot probe
634-634-002
36 inch pitot probe
634-634-003
Optional Accessories
Telescoping pitot probe
634-634-004
Static pressure probes
361-010-000
Duct plugs (100 pieces)
321-800-001
Rubber hose (8 ft.)
372-000-000
Carrying
case
534-550-001
HVAC conversion calculator
118-550-001
Specifications are subject to change without notice.