Chapter 4
Symptom
Possible Causes
Corrective Action
Flow reading
fluctuates badly
The flow is fluctuating
Use a longer time
constant.
“
ERR
” appears in
small letter and “dir
↑↓
” blinks along
with beeping
Flow direction changed
while taking sample
Make sure flow stays in
one direction (supply or
return) during sample.
Also, A
CCU
B
ALANCE
must remain in place for
one entire time constant
during a sample.
Movement of the
A
CCU
B
ALANCE
or
removal from its
measurement position
too soon will cause
errors.
"
ERR1
" appears
Temperature
compensation sensor is
damaged
Unit is or was recently
in ambient temperature
outside of operating
temperature
Return to the factory for
service.
Allow unit to stabilize in
the operating
temperature range of 32
- 140ºF (0 - 60ºC).
"
ERR2
" appears
The flow sensor is
giving an erroneous
reading
Return to the factory for
service.
"
ERR3
" appears
The A
CCU
B
ALANCE
Plus
has detected a
calibration fault
Return to the factory for
service.
“ERR4”
appears
Lithium Ion battery is
low or dead
Return to factory for
service.
“ERR5”
appears
Data RAM is
inoperative
Return to factory for
service.